Liquicap M FTI51, FTI52
Installation
Hauser
31
3.4
Installation examples
3.4.1
Rod probes
Conductive tanks (metal tanks)
If the process connection of the probe is insulated from the metal tank (e.g. using seal material),
connect the ground connection on the probe housing to the tank using a short line.
!
Note!
• A fully insulated rod probe may be neither shortened nor extended.
• If the insulation of the probe rod is damaged, this will falsify the measurement result.
• These application examples show vertical installation for MAX level limit detection.
FTI51 rod probe
L00-FTI5xxxx-11-06-xx-xx-004
FTI51 rod probe with ground tube
Nonconductive tanks (plastic tanks)
For installation in a plastic tank, use a probe with ground tube.
L00-FTI5xxxx-11-06-xx-xx-005