Electrical connection
Proline 400 HART
22
Hauser
5. Insert dummy plugs (corresponding to the housing degree of protection) into unused
cable entries.
NOTICE
Standard dummy plugs used for transportation do not have the appropriate degree of
protection and can result in damage to the device!
‣
Use suitable dummy plugs corresponding to the degree of protection.
5.6
Post-connection check
Are cables or the device undamaged (visual inspection)?
Do the cables comply with the requirements → 12?
Do the cables have adequate strain relief?
Does the supply voltage match the specifications on the transmitter nameplate ?
Is the terminal assignment correct → 13?
If supply voltage is present, do values appear on the display module?
Are all housing covers installed and the screws tightened with the correct tightening torque?