Proline 400 HART
Mounting
Hauser
11
4.3
Transmitter post-installation check
The post-installation check must always be performed after the following tasks:
• Turning the transmitter housing
• Turning the display module
Is the device undamaged (visual inspection)?
Turning the transmitter housing:
• Is the fixing screw firmly tightened?
• Is the connection compartment cover screwed on tightly?
• Is the securing clamp firmly tightened?
Turning the display module:
• Is the connection compartment cover screwed on tightly?
• Is the securing clamp firmly tightened?