Commissioning
Gammapilot M FMG50
32
Hauser
1
2
Alarm
on
of
f
Max
SW
Zero/
I
Span/
II
Display
3
4
A0039285
1
Operating key for empty calibration (function I)
2
Operating key for full calibration (function II)
3
DIP switch for alarm current (SW-defined/Min. alarm)
4
DIP switch for locking and unlocking the measuring device
•
Empty calibration:
Press and hold the operating key for empty calibration (I) > 3 s
•
Full calibration:
Press and hold the operating key for full calibration (II) > 3 s
•
Background calibration:
Simultaneously press and hold the operating key for empty
calibration (I) and the operating key for full calibration (II) > 3 s
•
Reset to factory defaults:
Simultaneously press and hold the operating key for empty
calibration (I) and full calibration (II) > 12 s. The LED starts flashing. When the flashing
stops, the device is reset to the factory default settings.
A reset deletes all calibrations!
Operation via keys is only active if a display is not attached
6.4.1
Status and power LEDs
A green LED that signals the supply, status and button activation feedback is provided on the
electronic insert.
Behavior of the LED
• When the measuring device is started, the LED flashes slowly until the current output
corresponds to the measured value
• After commissioning, the LED is lit constantly provided that the main device is switched on
and a local display is not attached
• When a key is pressed, the LED flashes to confirm the key activation
• When a reset is performed, the LED flashes as long as both keys are pressed and the reset is
not yet active (countdown). The LED stops flashing once the reset is active.
This LED is only active if a display is not attached