42
Hauser
Operation
FQG61/FQG62; order code
020, option A
Switching ON the radiation
L
CAUTION
Safety instructions for switching ON the radiation
‣
Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the
product vessel).
‣
The radiation may only be switched ON by properly instructed staff.
Switching ON the radiation
The source container is in the "OFF" position.
1. Press the cover hard against the source container and turn the cover approx. 45°
counterclockwise as far as it will go
2. Remove the cover
1.
2.
A0018415
3. Using the key, turn the closing cylinder approx. 45° in the counterclockwise direction
3.
A0033938
4. Pull out the lock until the limit stop.
5. Only for source containers with order code 670 "Additional function", option WA "Density
measurement > Fixation ON": release setscrew using Allen key.
6.
L
WARNING
Rotating the insert over the locking pin leads to the removal position of the source insert.
‣
Do not press the sealed locking pin (A)
Turn the insert 180° in the counterclockwise direction
A
5 mm
5.
6.
4.
A0033939
7. Press in the cylinder lock with the key and turn it by approx. 45° in the clockwise direction