CYR52
Electrical connection
Hauser
23
Connect the ultrasonic cleaning system in the following order:
1.
Connect the control cable (not included in the delivery, H03VV-F 2x0.75 is recommended)
to terminals 1 and 2 at the ultrasonic generator, and to a free relay at the transmitter at
terminals 41 and 42. The assignment of the individual wires is not relevant here. The
maximum cable length is 3 m (9.84 ft). Make sure that the control cable is strain-relieved
in both devices.
2.
Connect the cable secured to the ultrasonic transducer to the terminal block of the
ultrasonic generator as follows: connect the yellow-green protective conductor first. Then
connect the gray cable to the negative end (minus) and then the pink cable to the positive
end (plus). Ensure strain relief in the ultrasonic generator.
3.
Connect the mains connection cable (not included in the delivery, H05VV-F3 G0.75 is
recommended) to the appropriate terminals of the ultrasonic generator (
item 3). Ensure strain relief in the ultrasonic generator.
4.2
Post-connection check
Carry out the following checks after electrical connection:
Device condition and specifications
Notes
Are the devices and cables free from damage on the outside?
Visual inspection
Does the mains voltage match the specifications on the nameplate?
Electrical connection
Notes
Are the mounted cables strain-relieved?
Cable runs without loops and cross-overs?
Are the cables correctly connected according to the wiring diagram?
Are all the screw terminals firmly tightened?
Содержание CYR52
Страница 12: ...Installation CYR52 12 Endress Hauser Mounted on a pipe a0022757 Fig 5 Ultrasonic transducer mounted on a pipe ...
Страница 14: ...Installation CYR52 14 Endress Hauser a0022753 Fig 7 Dimensions in mm inch for assembly of CUS31 ...
Страница 15: ...CYR52 Installation Endress Hauser 15 a0022756 Fig 8 Dimensions in mm inch for pipe mounting ...
Страница 32: ...www addresses endress com 71261315 ...