CYR52
Installation
Hauser
17
3.3
Mounting the ultrasonic transducer
The ultrasonic transducer is secured to assemblies with two brackets that are supplied, and to
pipes with two worm drive hose clips that are supplied.
a0022759
Fig. 10: Ultrasonic contact paste(1)
Mount the ultrasonic transducer as follows:
1.
Using the cleaning cloth (scope of delivery) clean the contact surface of the ultrasonic
transducer and the contact surface of the assembly or pipe.
2.
Coat the contact surface of the ultrasonic transducer (
10, item 1) with a 5 mm
(0.2")stripe ultrasonic contact paste.
3.
Mount the ultrasonic transducer face to face with the turbidity sensor.
4.
In the case of pipes:
Tighten the two worm drive hose clips with a torque of 2.5 Nm (1.85 lbf ft).
In the case of assemblies:
Tighten the nuts of the brackets with a torque of 2 Nm (1.47 lbf ft).