Access to the seals
1
2
3
4
A0038663
12
Seals
1
Viton O-ring, retraction pipe
2
Viton O-ring, between ball valve and bottom part of bayonet lock
3
Viton O-rings, bottom part of bayonet lock
4
Viton O-ring, sensor holder
1. Only when replacing the O-ring, item 2: unscrew the vent cock (with safety bracket).
2. Only when replacing the O-ring, item 2: using a hook spanner, unscrew the bottom
part of the bayonet lock.
The seals can now be accessed.
Replacing the seals and assembling the assembly
1. Lightly grease the O-rings (e.g. with Syntheso Glep 1).
2. Replace the seals (O-rings) if necessary.
3. If you have not already done so, wrap Teflon tape around the thread of the retraction
pipe.
This protects the seals from damage when inserting the retraction pipe.
4. Grease the retraction pipe.
5. Reassemble the assembly.
6. Make sure that the locking ring above the bayonet nut is correctly seated.
7. Remove the Teflon tape once you have inserted the retraction pipe.
8. Check for leaks before returning the assembly to the measuring position.