Installation
CleanFit CPA 451
12
Hauser
7.
Screw the ring nut on the sensor (Fig. 15) and the sensor firmly into the internal
thread of the sensor holder.
8.
Press the sensor holder back into the outer sleeve and screw the bayonet nut
clockwise.
9.
Plug the locking pin into the drill holes of the bayonet nut and the sensor holder.
10. Open the ball valve and press the outer leeve as far as possible in the direction of
the process. This is easier when you lubricate the outer leeve. Please refer to the
notes of the chapter "Maintenance".
11. Lock the bayonet joint and secure the sensor holder using the fastening screws.
12. You can change the position of the upper assembly part, e.g. for another position
of the handles (due to the installation situation).
Loosen the retaining screws (Fig. 16) and turn the complete upper assembly part
around its own axis to the desired position.
13. Lock this position by tightening the 2 screws until the assembly part cannot turn any
more.
C07-CPA451xx-11-07-06-xx-005.eps
Fig. 15:
9
Sensor holder with
internal thread
10
Sensor
11
Ring nut
C07-CXA451xx-11-07-06-xx-008.eps
Fig. 16:
14
Retaining screw(s)
(2 screws, only one is visible in the figure)