Operation
Cerabar M, Deltapilot M
34
Hauser
6
Operation
6.1
Operation methods
6.1.1
Operation without an operating menu
Operation methods
Explanation
Graphic
Description
Local operation without
device display
The device is operated using the
operating keys on the electronic
insert.
Display
Zero
Span
A0045577
6.1.2
Operation with an operating menu
Operation with an operating menu is based on an operation concept with "user roles"
Operation methods
Explanation
Graphic
Description
Local operation with
device display
The device is operated using the
operating keys on the device dis-
play.
E
+
–
A0029999
Remote operation via
FieldCare
The device is operated using the
FieldCare operating tool.
A0030002
6.1.3
IO-Link
IO-Link information
IO-Link Smart Sensor Profile 2nd Edition
Supports
• Identification
• Diagnosis
• Digital Measuring Sensor (as per SSP 4.3.3)
IO-Link is a point-to-point connection for communication between the measuring device
and an IO-Link master. The measuring device features an IO-Link communication interface
type 2 (pin 4) with a second IO function on pin 2. This requires an IO-Link-compatible
assembly (IO-Link master) for operation. The IO-Link communication interface enables
direct access to the process and diagnostic data. It also provides the option of configuring
the measuring device while in operation.