Technical data
CCS120D
36
Hauser
12.2 Performance characteristics
12.2.1 Reference operating conditions
Temperature
30 °C (86 °F)
pH value
pH 7.2
12.2.2 Response time
T
90
approx. 60 s (with increasing and decreasing concentration)
12.2.3 Measured value resolution of sensor
0.01 mg/l (ppm)
12.2.4 Maximum measured error
±2 % or 200 µg/l (ppb) of value measured (depending on which value is higher)
LOD (limit of detection)
LOQ (limit of quantification)
0.022 mg/l (ppm)
0.072 mg/l (ppm)
1)
Based on ISO 15839. The measured error includes all the uncertainties of the sensor and transmitter (electrode
system). It does not contain all the uncertainties caused by the reference material and adjustments that may have
been performed.
12.2.5 Repeatability
0.008 mg/l (ppm)
12.2.6 Nominal slope
4 nA per 1 mg/l (ppm) (under reference operating conditions)
12.2.7 Long-term drift
< ±3 % per month
12.2.8 Polarization time
Initial commissioning
Up to 24 h
After replacing the membrane cap
Typically 1 to 6 h
Recommissioning
Pour approx. 4 to 24 h
12.2.9 Operating time of the electrolyte
3 to 6 months (depending on water quality)
12.2.10 Operating time of membrane cap
With electrolyte
Typically 3 to 6 months, depending on water quality
Without electrolyte
> 2 years (25 °C (77 °F))