Proline Prosonic Flow 92F
ATEX II3G
0
Hauser
19
Service connector
The service connector (for connection, see Fig. 3 to Fig. 7, e) is only used to connect service interfaces
approved by Hauser.
#
Warning!
The service connector may not be connected in a potentially explosive atmosphere.
Technical data
Dimensions
The dimensions of the Ex transmitter housing and the sensor correspond to the standard versions.
Please refer to the Technical Information for these dimensions.
!
Note!
Associated "Technical Information":
Proline Prosonic Flow 92F
TI073D/06/…
Terminals
1 (+)
2 (–)
or
1 (+)
2 (–)
Prosonic 92F**-***********H
PROFIBUS PA
PROFIBUS PA
Intrinsically safe circuit
EEx nL
EEx nL
Functional values
U
B
9 to 32 V DC
9 to 32 V DC
I
B
16 mA
16 mA
P
1 W
1 W
Safety related values
Ui
17.5 V
24 V
I
i
600 mA
250 mA
P
i
8.5 W
1.2 W
L
i
10
μ
F
10
μ
F
C
i
5 nF
5 nF
Terminals
1 (+)
2 (–)
or
1 (+)
2 (–)
Prosonic 92F**-***********K
FOUNDATION Fieldbus
FOUNDATION Fieldbus
Intrinsically safe circuit
EEx nL
EEx nL
Functional values
U
B
9 to 32 V DC
9 to 32 V DC
I
B
16 mA
16 mA
P
1 W
1 W
Safety related values
Ui
17.5 V
24 V
I
i
600 mA
250 mA
P
i
8.5 W
1.2 W
L
i
10
μ
F
10
μ
F
C
i
5 nF
5 nF