![Endress+Hauser ATEX II1/2GD Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/endress-hauser/atex-ii1-2gd/atex-ii1-2gd_safety-instructions_2407689029.webp)
Proline Promass 80
XA00097D
Hauser
29
FB = Full bore
)
)
The minimum
medium temperature
is –50 °C for Promass A/F/H/I/M/P/S, and –40 °C for Promass E.
Transmitter
The remote version transmitter has a T6 temperature class rating when installed in the Ex d housing for ope-
ration at ambient temperatures up to T
a
= 60 °C. The maximum ambient temperature range is –20 to +60 °C.
A version for
ambient temperatures
T
a
up to –20 °C is optionally available.
DN
[mm]
T
a
[°C]
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promass 8*I**-…
8…15
+60
60
95
130
150
150
150
15 FB, 25
25 FB
70
85
120
150
150
150
40, 40 FB
50, 50 FB
80
DN
[mm]
T
a
[°C]
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promass 8*M**-…
8...15
+60
55
95
130
150
150
150
25...40
60
95
130
150
150
150
50
65
95
130
150
150
150
80
65
80
110
150
150
150
DN
[mm]
T
a
[°C]
T6
(85 °C)
T5
(100 °C)
T4
(135 °C)
T3
(200 °C)
T2
(300 °C)
T1
(450 °C)
Promass 8*P**-…
Promass 8*S**-…
8
+50
–
65
100
160
200
200
15…25
50
75
115
180
200
200
40
55
75
115
180
200
200
8
+60
–
65
100
160
200
200
15…40
–
75
115
180
200
200
50
60
75
110
180
200
200
Содержание ATEX II1/2GD
Страница 20: ...0ATEX II2GD II1 2GD IECEx Zone 1 Zone 21 Zone 0 1 Proline Promass 80 20 Endress Hauser ...
Страница 21: ...Proline Promass 80 ATEX II2GD II1 2GD IECEx Zone 1 Zone 21 Zone 0 1 0 Endress Hauser 21 ...
Страница 49: ...Proline Promass 80 XA00097D Endress Hauser 49 ...
Страница 50: ...www endress com worldwide XA00097D 06 A3 14 12 71146622 FM SGML 9 0 ...