Operation
PlasmaQuant MS product family
54
Fig. 25 Instrument summary window
5.3
Starting the vacuum system
If the system does not reach the required vacuum pressure within 10 minutes, an error
will be displayed and the vacuum system will be shut down.
1.
Click the [V
ACUUM
] button on the Main Toolbar of the ASpect MS Software, or
choose V
ACUUM
O
N
from the A
CTIONS
menu, or press [SHIFT+F6]. A sequence will
automatically initiate the appropriate vacuum components to bring the instrument
to the required system pressure.
2.
Watch the Status Bar (at the bottom of the Main window) to see the progress.
Countdown times (in seconds) for the various stages are displayed.
3.
After restarting the vacuum system it is recommended to carry out an Automated
Instrument Setup (see section "Instrument Automated Setup" on page 59).