PlasmaQuant MS product family
Installation and commissioning
43
4.3.1
Installation of the torch
CAUTION
Danger of injury from breaking glass! Always take care when installing or removing the
parts.
1.
Fit one quick-connect fitting to each length of tubing.
2.
Fit the quick-connect snap fitting to the other end of one length
of tubing. Slide the fittings onto the torch. The line with the
quick-connect fitting on the other end goes on the side arm
nearest the spherical joint. This will be the auxiliary gas line
(2)
.
Tubing
(1)
will be the plasma gas line.
Torch tubing: 4.75 mm ID / 1.52 mm wall
length 300 mm
3.
Carefully insert the ignition lead into the open end of the other
length of tubing
(1)
until only the connector is exposed.
4.
Open the plasma compartment lid on the front side of the
instrument.
5.
Open the clamp inside the torch compartment. Place the torch in
the clamp sliding it forward between the induction coils.
6.
To position the torch correctly, slide the torch forward until the
glass nubs on its bottom side are touching the outside of the
clamp.
Installation of the torch