Transport and storage
multi X 2500
116
2.
Remove the combustion tube.
3.
Detach the gas transfer line from the connections and remove it.
4.
Remove the auto-protection assembly.
5.
Remove the combustion furnace (
→
"Removing the combustion furnace", p. 116).
6.
Detach all hoses or put the hose ends into small bags and attach them to the basic
device with adhesive tape.
7.
Secure the top cover of the basic device and the maintenance flaps on the right-
hand side of the device with adhesive tape.
1.
Detach the hose connection to the measuring cell from the sulfuric acid container
with outlet. Remove the safety attachment. Remove the empty sulfuric acid
container and rinse it.
2.
Remove the storage container for the sensor electrode and rinse it.
3.
Detach the electrical connection of the electrodes. Put all electrodes into their
original packaging. When doing this, observe the servicing and maintenance
instructions for the electrodes (
→
"Clean and store the electrode", p. 105).
4.
Remove the measuring cell and rinse it.
WARNING! The electrolyte solution of the measuring cell "high sensitive" contains
concentrated acetic acid. Be careful when emptying the measuring cell!
5.
Attach the extraction hose to the chlorine module using adhesive tape.
12.3
Removing and installing the combustion furnace
12.3.1
Removing the combustion furnace
WARNING
Danger of electric shock! Before removing the combustion furnace switch off the
multi X 2500 using the mains switch and disconnect the mains plug from the power
outlet!
CAUTION
There is a risk of burns on the combustion furnace! Only remove the combustion
furnace when the device is cold or allow the device to cool down sufficiently!
1.
Exit the multiWin software. Allow the basic device to cool down.
2.
Switch off the multi X 2500 and the chlorine module using the mains switch and
pull the mains plugs from the power outlet. Cut the gas supply.
3.
Remove the top cover. Open the front doors and remove them from the device.
4.
Remove the combustion tube (
→
"Removing and cleaning the open combustion
tube", p. 89 or "Removing and cleaning the multi-purpose combustion tube",
p. 100).
In the chlorine module