Maintenance and care
multi X 2500
100
10.4.4
Removing and cleaning the multi-purpose combustion tube
WARNING
Danger of explosion! The hose connections for argon and oxygen on the combustion
tube must not be mixed up when reinstalling the assembly.
CAUTION
Risk of burns! Only remove the combustion tube when the device is cold!
NOTICE
Alkaline salts (hand perspiration) cause crystallizations in the quartz glass when
heating the combustion furnace which reduce the service life of the combustion tube.
Do not touch the combustions tube with your bare hands during reinstallation. Wear
protective gloves. If necessary, clean the combustion tube externally before installing it
in the combustion furnace (e.g. by wiping it with moist cellulose).
NOTICE
Closing the seal when the combustion tube has not been inserted into the furnace may
destroy the pneumatic seal of the auto-protection assembly. Only move the toggle
switch down if a combustion tube has been inserted. Alternatively, we recommend to
cut the gas supply during maintenance.
1.
Disconnect the gas transfer line from the sulfuric acid container with outlet in the
chlorine module.
2.
Exit the multiWin control software and allow the analyzer to cool down.
3.
Switch off the multi X 2500 and the chlorine module. Remove the mains plug from
the power outlet. Switch off the gas supply.
4.
In vertical operating mode, remove the cover at the top.
5.
Open the front doors of the analyzer.
6.
Open the pneumatic seal to the auto-protection assembly. To
do this, move the toggle switch up. This releases the
combustion tube.
7.
Allow the device to cool down.