multi EA 5100
Maintenance and care
159
14.14.1 Replacing the water traps
81
1
2
3
4
5
6
Figure 76Replace the water traps
1 FAST connector (angled)
2 Screw connection
3 Disposable retention filter (water
trap)
4 Clamp
5 Aerosol trap (large water trap)
6 Hose connector (to the cooling
block)
ð
Replace the water traps at least every 6 months.
}
Remove the hose connections from the water traps. Remove the water traps from
the clamps.
}
Assembly the new water traps:
The "INLET" marking on the large water trap (aerosol trap) must be facing down-
ward.
The red labeling on the small water trap must face upward.
}
Insert the new water traps into the clamps. The large water trap has to be positioned
at the bottom.
}
Connect the hose connections to the water traps.
NOTICE! For angled FAST connectors: Do not push the connections too far into the
limbs of the FAST connectors. The gas flow may be impeded otherwise.
}
Check the system for leaks.
ü
The detection module is ready for operation again.
14.14.2 Replacing the halogen trap
CAUTION
Skin and respiratory system irritation due to quartz wool
Quartz wool tends to form dust. Irritation can occur after breathing in or skin contact
with this dust.
¡
Avoid the formation of dust when working with quartz wool.
¡
Wear protective clothing and gloves.
¡
Work under an extractor or wear a respiratory mask.