9.
Place the filter unit into the analyzer and tighten the base with the
four screws.
10. Check upstream of membrane for liquid contamination and clean and
dry out before opening the sample supply valve.
11. Check connections for gas leaks. Using a liquid leak detector is recom
-
mended.
Replacing the scrubber and scrubber efficiency indicator
1.
Close the sample supply shut-off valve. Allow all residual gas to dissi
-
pate as indicated by no flow on the sample bypass flow meter.
2.
Unscrew the compression nuts on the inlet end of the scrubber and
scrubber efficiency indicator assembly.
3.
To install the new scrubber and indicator, insert the inlet and outlet
tubes into the compression fittings of a new scrubber and scrubber
efficiency indicator assembly, ensuring each are oriented correctly, as
shown below.
4.
Tighten all new fittings 1-1/4 turns with a wrench from finger tight.
For connections with previously swaged ferrules, thread the nut to the
previously pulled up position, then tighten slightly with a wrench.
5.
Reset the scrubber lifetime monitor with the
New Scrub Installed
parameter and the
General Fault Alarm
with the Reset option for the
General Alarm
DO
parameter (see
“To change parameters in Mode
2”
in the Description of Device Parameters for your analyzer).
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.