Condumax CLS15D/16D/21D
Installation
Hauser
13
2
1
3
35
.38
(1
)
A0019015
1
With NPT ½" thread in T-piece or cross
fitting
1
T-piece or cross fitting (DN 32, 40 or 50)
2
Glue-in VC threaded coupling (NPT ½" for
DN 20)
3
Glue-in adapter coupling (for DN 32, 40, 50)
NPT ½”
160 (6.30)
Ø 29 (1.14)
Ø 6 (0.24)
30 (1.18)
25 (0.98)
Ø 6 (0.24)
38.5
(1.52)
1
2
3
A0019014
2
With NPT ½" thread in flow assembly
71042405, dimensions in mm (inch)
1
Sensor holder NPT ½"
2
Inlet
3
Outlet
Ensure that the electrodes are fully immersed in the medium during measurement. The
immersion depth must be at least 35 mm (1.38"). If the sensor is being used in the
ultrapure water range, you must work under air-evacuated conditions. Otherwise the CO
2
in the air can dissolve in the water and its (weak) dissociation can increase the
conductivity by up to 3 µS/cm.