5.2 

Connecting the sensor

The electrical connection of the sensor to the transmitter is performed using the measuring

cable CYK10.

GN/YE

YE

GN

BN

WH

GND

+

Com A

Com B

GY

  A0024019

 10

Measuring cable CYK10

NOTICE

Mechanical twist protection for CLS15D and CLS21D

If too much force is applied to the Memosens head, this can shear the connections and

therefore destroy the sensor

There is no need to exert excessive force when connecting the sensor to the cable coupling.

Always connect the sensor gently.

If the Memosens coupling clearly will not close, check the coupling for dirt or mechanical

damage and make sure you are turning in the right direction. Pay attention to the lock

symbol on the coupling.

Use another Memosens cable if necessary.

5.3 

Ensuring the degree of protection

Only the mechanical and electrical connections which are described in these instructions and

which are necessary for the required, designated use, may be carried out on the device

delivered.

Exercise care when carrying out the work.

Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC

interference immunity) agreed for this product can no longer be guaranteed due, for example,

to covers being left off or cable (ends) which are loose or insufficiently secured.

5.4 

Post-connection check

Device condition and specifications

Notes

Are the outside of the sensor, assembly, cable undamaged?

Visual inspection

Electrical connection

Notes

Are the installed cables strain-relieved and not twisted?

Is a sufficient length of the cable cores stripped, and is it positioned
in the terminal correctly?

Check the fit (by pulling gently)

Are all the screws terminals properly tightened?

Tighten

Содержание 16D/21D

Страница 1: ...Products Solutions Services Operating Instructions Condumax CLS15D 16D 21D Sensors with Memosens protocol For conductive measurement of conductivity in liquids 0 BA01147C 07 EN 03 16 71333350...

Страница 2: ......

Страница 3: ...ry 10 3 4 Certificates and approvals 11 4 Installation 12 4 1 Mounting the sensor 12 4 2 Post installation check 16 5 Electrical connection 16 5 1 Connection conditions 17 5 2 Connecting the sensor 19...

Страница 4: ...ngerous situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences Consequences of non compliance if applicable Corrective action This symb...

Страница 5: ...liquids They are used in the following fields Sensor Applications Hazardous areas Condumax CLS15D Measurements in pure and ultrapure water Approved for Ex zone 0 Condumax CLS16D Measurements in pure...

Страница 6: ...this document The certified conductivity sensors CLS15D CLS16D CLS21D may only be connected via the measuring cable CYK10 G I to the certified intrinsically safe digital sensor output circuits of mea...

Страница 7: ...to the certified intrinsically safe digital sensor output circuits of measuring device Liquiline M CM42 KV The electrical connection must be made according to the wiring diagram of the transmitter Me...

Страница 8: ...5 C T4 5 C Ta 65 C T6 II 3G Condumax CLS21D V 20 C Ta 135 C T3 20 C Ta 115 C T4 20 C Ta 65 C T6 II 3G If the specified medium temperatures are complied with temperatures that are not permitted for the...

Страница 9: ...protected against impact and moisture The original packaging offers the best protection The permitted ambient conditions must be observed see Technical data If you have any questions please contact yo...

Страница 10: ...he order code The order code and serial number of your product can be found in the following locations On the nameplate In the delivery papers Obtaining information on the product 1 Go to the product...

Страница 11: ...n the plug in head 3 4 3 EHEDG CLS16D only Validated as follows Cleanability in accordance with EHEDG Document 2 Sterilizability in accordance with EHEDG Document 5 Bacteria tightness in accordance wi...

Страница 12: ...NPT or or clamp 1 As an option the sensor can also be installed using a commercially available T piece or cross fitting or using a flow assembly NOTICE Incorrect mounting or disassembly The Memosens h...

Страница 13: ...98 6 0 24 38 5 1 52 1 2 3 A0019014 2 With NPT thread in flow assembly 71042405 dimensions in mm inch 1 Sensor holder NPT 2 Inlet 3 Outlet Ensure that the electrodes are fully immersed in the medium d...

Страница 14: ...must work under air evacuated conditions Otherwise the CO2 in the air can dissolve in the water and its weak dissociation can increase the conductivity by up to 3 S cm 4 1 3 CLS21D Clamp connection Bo...

Страница 15: ...stalled via a flow assembly NOTICE Incorrect mounting or disassembly The Memosens head could become loose and fall off resulting in total sensor failure Only mount the sensor via the process connectio...

Страница 16: ...Post installation check Are the sensor and cable undamaged Is the sensor installed in the process connection and is not suspended from the cable 5 Electrical connection LWARNING Device is live Incorre...

Страница 17: ...ensors for zone 0 IIC 2G CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D IIC 1G IIC 3G EX Zone 1 Zone 0 Zone 2 CYK10 Ex ia Ga CPS41D CPS42D CYK10 Ex ia Ga EX CLS50D CLS15D CLS16...

Страница 18: ...zone 2 CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D EX CYK10 Ex ic CPS41D CPS42D CYK10 Ex ic EX CLS50D CLS15D CLS16D CLS21D COS22D COS51D CPS471D CPS491D CPS441D CLS82D CPS3...

Страница 19: ...er Memosens cable if necessary 5 3 Ensuring the degree of protection Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required d...

Страница 20: ...cility make sure the connections are correctly fitted Do not install the assembly in the process if you cannot make the correct connection reliably 7 Maintenance LCAUTION Corrosive chemicals Danger of...

Страница 21: ...desulfurization or sewage treatment plants Use a mixture of hydrochloric acid 3 and thiocarbamide commercially available and then rinse thoroughly with plenty of clear water 4 Buildup containing prot...

Страница 22: ...ice for information on replacing the seal and recalibration in the factory 8 2 Return The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered...

Страница 23: ...on to water at 25 C 77 F CLS15D A 0 04 to 20 S cm CLS15D B L 0 10 to 200 S cm CLS16D 0 04 to 500 S cm CLS21D 10 S cm to 20 mS cm Temperature CLS15D 20 to 100 C 4 to 212 F CLS16D 5 to 100 C 23 to 212 F...

Страница 24: ...rence measuring system traceable to NIST or PTB The exact cell constant is entered into the quality certificate supplied The uncertainty of measurement in determining the cell constant is 1 0 CLS21D E...

Страница 25: ...mal operation 5 to 120 C 23 to 248 F Sterilization max 45 min Max 150 C 302 F at 6 bar 87 psi absolute CLS21D 20 to 135 C 4 to 275 F at 3 5 bar 50 psi absolute 1 Threaded versions max 30 minutes The m...

Страница 26: ...20 40 60 80 100 120 140 C 212 p abs T A0031428 EN 11 Mechanical pressure temperature resistance A Can be sterilized for a short time 1 hour CLS16D 6 5 20 60 100 120 150 23 60 68 120 180 240 248 300 20...

Страница 27: ...66 lbs depending on version CLS16D Approx 0 13 to 0 75 kg 0 29 to 1 65 lbs depending on version 9 5 2 Materials CLS15D Electrodes Polished stainless steel 1 4435 AISI 316L Sensor shaft Polyethersulfo...

Страница 28: ...1 1 2 as per ISO 2852 also suitable for TRI CLAMP DIN 32676 Tuchenhagen VARIVENT N DN 50 to 125 NEUMO BioControl D50 CLS21D Thread G1 NPT 1 thread Clamp 2 as per ISO 2852 Sanitary connection DN 25 as...

Страница 29: ...Condumax CLS15D 16D 21D EU Declaration of conformity Endress Hauser 29 10 EU Declaration of conformity 10 1 ATEX NEPSI II 1G Ex ia IIC T3 T4 T6 Ga IECEx Ex ia IIC T3 T4 T6 Ga...

Страница 30: ...EU Declaration of conformity Condumax CLS15D 16D 21D 30 Endress Hauser 10 2 ATEX NEPSI II 3G Ex ic IIC T3 T4 T6 Gc...

Страница 31: ......

Страница 32: ...www addresses endress com 71333350 71333350...

Отзывы: