accuracy  of  subsequent  measurements  is  in  question.  If  possible,  an  accelerometer  that  has 
experienced a large ZMO shift should be replaced or returned for recalibration. Any ZMO reading 
approaching  one-half  of  the  excitation  voltage  up  to  the  excitation  voltage  is  a  sure  sign  of 
accelerometer failure. The simple resistance check described in the “Initial Checkout” section above 
can also be used to evaluate accelerometer health. 

 

After  performing  the  above  tests,  if  you  are  still  uncertain  about  the  health  of  your  accelerometer 
please contact Applications Engineering. An Application Engineer will be able to discuss the health 
of  your  accelerometer  in  more  detail  and  suggest  if  it  is  healthy,  needs  recalibration  or  needs 
replacement. Endevco maintains an accelerometer recalibration service with NIST traceability in the 
United  States.  Endevco  European  Regional  offices  maintain  accelerometer  recalibration  services 
with traceability to applicable national standards. 

 

QUESTIONS

 

 

If  you  have  any  questions  regarding  the  use  of  these  accelerometers  (or  any  other  Endevco

®

 

product) please contact Endevco’s Application Engineering at +1 (866) ENDEVCO  [+1 (866) 363-
3826] in North America, or your local sales representative. 
 

Содержание 727

Страница 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Страница 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Страница 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Страница 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Страница 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Страница 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Страница 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Страница 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Страница 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Отзывы: