ENDEVCO 727 Скачать руководство пользователя страница 3

IM727 Rev A, Page 3 of 9 

 

 

®

 

HANDLING PRECAUTIONS

 

 

Although  undamped  accelerometers  are  more  susceptible  to  damage  by  handling,  the  follow 
precautions should be taken when handling any accelerometer, whether it is damped or undamped. 
The  amplification  factor  (or  Q)  of  the  undamped  accelerometer  at  the  resonant  frequency  of  its 
seismic  element  can  approach  100  (10  for  the  damped  accelerometers).  This  means  that  if  an 
unmounted  unit  were  dropped  on  a  metal  surface  (such  as  a  table  or  bench  top)  causing  it  to 
experience a shock input of 2,000 g, the high-frequency content of the metal-to-metal impact could 
excite the resonance frequency,  resulting in  an  accelerometer  response  of  200,000  g (20,000  g  for 
the  damped  accelerometers).  It  has  been  shown  that  the  accelerometers  are  more  vulnerable  to 
damage  in  the unmounted state, as opposed to being mounted to a test subject of large mass. For this 
reason the unmounted unit must be handled very carefully to avoid such fast rise time impacts. 

 

Endevco®  drop-test  accelerometers  are  shipped  in  electrostatic  discharge  (ESD)  safe  packaging. 
When  handling  any  accelerometer  it  is  always  best  practice  to  handle  with  ESD  in  mind.  The 
accelerometer  should  only  be  handled  by  properly  grounded  technicians  (via  wrist  straps  or  heel 
straps) at ESD safe work stations. If ESD damage does occur it will typically result in a large shift in 
the  zero  measurand  output  (ZMO).  If  ESD  damage  is  sufficiently  high,  complete  accelerometer 
failure is likely, showing up as an extremely large ZMO or an open leg of the Wheatstone bridge. 

 

INITIAL CHECKOUT

 

 

Upon receipt, the accelerometer should be checked to ensure that it was not damaged in transit. A 
simple  resistance  test  is  a  quick  way  of  verifying  that  all  legs  in  the  Wheatstone  bridge  sensing 
element are intact. 

 

Resistance  Test  –  Open  the  accelerometer  container  and  unwind  a  few  inches  of  cable. 
Measure  the  input  resistance  (red  to  black)  and  output  resistance  (green  to  white  while 
shorting red and black) with an ohmmeter and a pair of clips leads. The measured resistance 
should be within the specified tolerance as listed on the product specification data sheet. 

 

Typical specification for Model 727 input/output resistance is 650Ω ± 

300Ω  

If any of the above resistance measurements are not within the noted specification there may be a 
problem with the accelerometer and the factory should be consulted for further troubleshooting. 

 

MOUNTING SURFACE PREPARATION AND INSTALLATION 
 

Special  care  should  be  taken  to  provide  a  smooth,  clean  mounting  surface  to  ensure  maximum 
transmissibility  of  the  shock  input.  Any  small  particles  or  debris  trapped  between  the  mounting 
surface  and  the  accelerometer  will  degrade  the  transmission  of  the  high- frequency  components  of 
the  shock  input.  An  uneven  mounting  surface  could  also  preload  the  accelerometer  and  cause 

Содержание 727

Страница 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Страница 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Страница 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Страница 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Страница 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Страница 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Страница 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Страница 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Страница 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Отзывы: