ENDEVCO 727 Скачать руководство пользователя страница 8

IM727 Rev A, Page 8 of 9 

 

 

®

Nyquist  frequency).  Once  digitized,  the  aliased  frequencies  will  fold  back  over  the  Nyquist 
frequency into the frequency range of interest resulting in errors in the data. To prevent aliasing, the 
sampling rate should be at least two times (and preferably ten times) the highest frequency present. 
The  highest  frequency  present  in  almost  all  situations  will  be  the  resonant  frequency  of  the  high 
shock  accelerometer.  Better  yet,  an  analog  anti-aliasing  filter  should  be  placed  before 
digitization  with  a cutoff  frequency  of  at  least  one-half  (and preferably  one-tenth)  the  sampling 
rate.  The  anti-aliasing  filter  safeguards  against  any  unpredictable  out-of-band  energy  and  is  the 
most reliable approach to acquiring good quality, non-aliased data. 

 

See  Endevco  Technical  Paper  TP284  for  further  information  on  high-frequency,  high-level  testing 
and calibration. 
 
 

REMOVAL OF THE ACCELEROMETER 

 
To  remove  the  727  accelerometer  from  your  application,  please  use  the  supplied  removal  tool, 
Endevco®  part  number  42894.  The  tool  should  be  placed  over  the  accelerometer  with  the  cable 
positioned  away  from  the  tool  to  shear  the  accelerometer  off  the  surface  of  your  application  by 
gently wriggling the tool in clockwise and counterclockwise directions. 
 
 

 

 

 

CLEANING 

 

If  desired,  dirty  accelerometers  may  be  wiped  clean  using  a  damp  cloth  and  a  solvent  such  as 
acetone. 

Do not soak or immerse 

the unit in any solvent or water. Do not use any sharp tool such 

as  a  screwdriver  to  remove  dirt  or  contaminants.  Any  temporary  adhesives,  such  as  wax  or 
cyanoacrylate, used to mount the accelerometer, should be cleaned with an appropriate solvent (such 
as acetone). 

 

RECALIBRATION

 

 

As   with   any   measuring   device,   optimum   accuracy   is   maintained   by   periodic   recalibration. 
Sensitivity and ZMO calibration should be performed at 6 to 12 month intervals depending on usage. 
Ordinarily,  recalibration  need  only  be  performed  at  12  month  intervals  if  it  is  known  that  the 
accelerometer  has  not  been  used  beyond  its  rated  specifications.  If  the  unit  is  used  under  severe 
environments,  it  may  be  desirable  to  use  shorter  calibration  intervals.  When  testing  at  high  levels 
(levels  exceeding  the  accelerometer’s  full-scale  range)  it  is  recommended  that  calibration  be 
performed every six months or after each high-amplitude test. 

 

The general health of the accelerometer can be assessed by measuring the ZMO and comparing the 
value  to  the  most  recent  calibration  certificate.  The  ZMO  can  be  checked  by  applying  a  well- 
regulated excitation voltage (10 Vdc in most cases, unless otherwise specified at time of order) to 
the red-black leads and reading out the differential voltage between the green-white leads. Changes 
in ZMO of up to ±20mV with time and usage are not uncommon and are not necessarily a sign of a 
damaged accelerometer. Changes of greater than ±20mV may indicate damage to the sensor and the 

Содержание 727

Страница 1: ...IM727 Rev A Page 1 of 9 INSTRUCTION MANUAL For Model 727 PIEZORESISTIVE DROP TEST ACCELEROMETERS IM727 Revision A 07 10 13 ...

Страница 2: ... addition to being undamped the 727 series has no mechanical stops maximizing both bandwidth and amplitude linearity Its undamped characteristic allows it to be used in applications where the peak acceleration is known and there is not a significant amount of out of band energy that could cause enough resonance excitation to exceed the overrange limit of the accelerometer Due to the severe environ...

Страница 3: ...does occur it will typically result in a large shift in the zero measurand output ZMO If ESD damage is sufficiently high complete accelerometer failure is likely showing up as an extremely large ZMO or an open leg of the Wheatstone bridge INITIAL CHECKOUT Upon receipt the accelerometer should be checked to ensure that it was not damaged in transit A simple resistance test is a quick way of verifyi...

Страница 4: ...HANDLING The 727 series of drop test accelerometers are supplied with integral cables which are terminated in pigtails The user can select an appropriate end connector There are two important considerations regarding the integral cable when installing the accelerometer First since the cable will be exposed to the same high level shocks as the accelerometer care needs to be taken to prevent damage ...

Страница 5: ...oresistive accelerometers including the 126 136 and 4430A Contact Applications Engineering to discuss your specific signal conditioning needs 2 Power and Signal Leads The cable leads are assigned as follows The cable shield is not connected to the accelerometer case It is recommended that the shield be connected to the power supply ground 3 Cable Length Considerations The 727 accelerometers are ca...

Страница 6: ...nput voltage attenuation discussion the resistance in the output leads will also cause attenuation of the output signal The reduced sensitivity is calculated as follows EO sensitivity into an infinite load from Cal Cert EOL loaded output sensitivity RO output bridge resistance from Cal Cert including cable RL load resistance RCO resistance of the spliced output wire c Cable Capacitance Long cable ...

Страница 7: ... to the Fourier transform Although resonance excitation cannot always be avoided it possible for the user to take measures to reduce the amount of high frequency energy the accelerometer will be subjected to BANDWIDTH AND AMPLITUDE LINEARITY Because the 727 series is undamped and does not use mechanical over travel stops accurate measurements are permitted over a wide frequency and amplitude range...

Страница 8: ...th and a solvent such as acetone Do not soak or immerse the unit in any solvent or water Do not use any sharp tool such as a screwdriver to remove dirt or contaminants Any temporary adhesives such as wax or cyanoacrylate used to mount the accelerometer should be cleaned with an appropriate solvent such as acetone RECALIBRATION As with any measuring device optimum accuracy is maintained by periodic...

Страница 9: ...out the health of your accelerometer please contact Applications Engineering An Application Engineer will be able to discuss the health of your accelerometer in more detail and suggest if it is healthy needs recalibration or needs replacement Endevco maintains an accelerometer recalibration service with NIST traceability in the United States Endevco European Regional offices maintain accelerometer...

Отзывы: