Equipment required:
•
28 volt power supply with an output mA meter
•
Oscilloscope
•
Digital multimeter
Follow the below steps:
1
Apply 28 VDC between pins A (+) and D and confirm that the current
is within the specifications in Table 4. Allow about >3 minutes for
warm-up
2
Install a 10k Ω, ¼ watt resistor between pins C & D.
3
Attach an oscilloscope to pins C and D.
4
Set the reference shaker frequency to 100Hz.
5
Observe a 100 Hz undistorted sine wave on the oscilloscope. If the
oscilloscope is DC coupled a DC bias of 2.50 volts (±3%)should also
be observed
Figure 9: A handheld shaker
(Endevco REF2520) is ideal for
testing system performance.
The shaker provides a one g
sine vibration level at selectable
frequencies. The accelerometer
under test is attached to the top
of the shaker.