Figure 25: The Fourier Transform plots vibration amplitude as a function of frequency. 

Render PSD 

Power  spectral  densities  (PSD)  are  used  to  characterize  random  vibration  signals.    A  PSD  is 
computed by multiplying each frequency bin in an FFT by its complex conjugate which results in 
the real only spectrum of amplitude in g

2

.  The key aspect of a PSD which makes it more useful 

than a FFT for random vibration analysis is that this amplitude value is then normalized to the 
frequency bin width to get units of g

2

/Hz.  By normalizing the result we reduce the dependency 

on  the  number  of  samples  in  the  original  acquisition.    The  area  under  the  curve  in  a  PSD 
represents  the  RMS  vibration  energy  in  that  frequency  range.  This  area  can  be  calculated 
precisely by exporting the PSD into a CSV file, summing the data in the desired frequency range, 
and multiplying by the frequency bin width, or the space between adjacent frequencies.  Slam 
Stick  Lab  filters  the  three  lowest  frequencies  to  0  to  remove  the  DC  component.    For  more 
information, visit ou

blog that covers FFTs, PSDs, and spectrograms

. 

Содержание Slam Stick C

Страница 1: ...User Manual Slam Stick User Manual Version No 2 0 1 User Manual Vibration Temperature Pressure Data Logger...

Страница 2: ...19 Slam Stick Lab Installation 20 Slam Stick Lab Configuration 21 General Configuration 23 Setting Triggers 26 Disabling Channels 30 User Calibration Configuration 30 Device Information 33 Update Fir...

Страница 3: ...e Buyer upon the return of the defective product Requests for compliance with this express limited warranty shall be honored only when made by the Buyer Refer to the Terms and Conditions at http www m...

Страница 4: ...s needed This software package allows the user to configure the recorder for different sample rates or triggering modes quickly review recorded vibration temperature and pressure data and export recor...

Страница 5: ...ate so that the user can perform detailed vibrational analysis A real time clock provides time stamps for all data recorded so that comparisons to real world events can be made Pressure and temperatur...

Страница 6: ...e of their versatility but they have two disadvantages For one they are AC coupled which means that it can t measure static accelerations like gravity and generally can t measure slow 2 Hz vibrations...

Страница 7: ...ent shock and vibration testing applications Refer to Mid s blog post on accelerometer selection for more information Table 1 provides an overview of the differences between the 3 products The catalog...

Страница 8: ...el Added Sensors Temperature Pressure Temperature Pressure Temperature Pressure Temperature Pressure Advantages Excellent value Same ease of use Piezoelectric accel Frequency 1 000 Hz Aluminum more ro...

Страница 9: ...ugh a variety of distributors including Amazon and Digi Key The included tape is used for calibration Calibration is not valid if other tape is used If modification to the mounting environment is poss...

Страница 10: ...dings and also to manually cycle through programmed triggers The coordinate system reference illustrates the axes directions of the device The tie down slot is included to use a zip tie or a string of...

Страница 11: ...is EAR99 The Harmonized tariff code or scheduled B number is 9031 80 8085 We accept all major credit cards as the preferred method of payment Purchase orders are accepted after approving credit refer...

Страница 12: ...t The section entitled Mounting Instructions provides detailed mounting instructions a The most common form of mounting the device is with the Mid recommended double sided tape provided with your kit...

Страница 13: ...at the yellow light is made by turning on both the green and red LED Battery cannot be charged due to extreme temperature Allow the device to reach the allowed charging temperature 0 C to 45 C If prob...

Страница 14: ...trigger Allow it to wait for the trigger or hold down button to skip to next trigger or recording Rapid blinking Green to Yellow Waiting for an acceleration trigger Allow it to wait for the trigger or...

Страница 15: ...age at room temperature with low humidity the Slam Stick has been tested in accordance with MIL STD 810G to ensure survival at extreme temperatures This testing included exposure to 40 C for eight hou...

Страница 16: ...Figure 5 The Slam Stick is charged by using the provided USB to Micro USB cable and will connect to any standard USB power supply CAUTION Be certain to insert the USB to Micro USB Cable into the micro...

Страница 17: ...harge will take one to two hours depending on the power source and the age of the battery NOTE The device may get warm during charging there is no reason for concern if you notice this The temperature...

Страница 18: ...t various sampling frequencies as well as expected recording time for a 2GB memory card when a Slam Stick is plugged in to an external power source The use of a consistent power source such as a compu...

Страница 19: ...econds after midnight meaning a file that started recording at 9 00am will be stored as SSX32400 because 9 00am is exactly 32 400 seconds after midnight NOTE If the folders are named with the incorrec...

Страница 20: ...ck devices will not initiate a new recording if the 4GB limit is reached If a data file becomes corrupted or does not display data i e a 0 Kb file in the Slam Stick Lab the Slam Stick may have stopped...

Страница 21: ...Stick to best meet your testing needs Figure 6 Video 1 of Mid s how to video series goes through an example configuration The video is available on YouTube here To open up the configuration settings...

Страница 22: ...w displays available Slam Sticks for configuration 4 Select OK The configuration settings window will appear allowing the user to adjust the measurement settings on the Slam Stick Figure 8 Configurati...

Страница 23: ...for the trigger condition it will remain off General Configuration The General configuration settings are the Slam Stick settings that appear when the configuration settings window is opened This is...

Страница 24: ...asure the necessary vibrations shock events expected during use As a guideline it is recommended that the sampling frequency be set to no less than ten 10 times the frequency of the vibration expected...

Страница 25: ...or this configuration it is recommended to use the default value on the Slam Stick by unchecking the box next to this configuration NOTE Frequencies higher than the anti aliasing cutoff frequency will...

Страница 26: ...offset between the user and the UTC Setting Triggers Located to the right of the General tab in the configuration settings window is the Triggers tab Triggers are configuration settings useful when a...

Страница 27: ...re multiple triggers are set Slam Stick will wait for each trigger setting to be accomplished before a recording begins But some triggers have priority in order they appear on the screen For example i...

Страница 28: ...UTC or the specific time zone of the user This trigger is useful if the user will not have access to the Slam Stick for an extended period of time before a recording is needed This trigger is also us...

Страница 29: ...the trigger Note that this is not a concern for Slam Stick S uses a piezoresistive accelerometer customers But if the user would prefer to use the piezoelectric accelerometer as the trigger you will...

Страница 30: ...se temperature and pressure data are required for acceleration measurements to be calibrated By removing the checkbox from a channel this deactivates that channel on the Slam Stick and no data will be...

Страница 31: ...up a trigger or view device information Select the device you wish to adjust the calibration for click OK and enter the configuration window as shown below 3 Click the Create User Calibration button a...

Страница 32: ...12 Figure 12 The calibration editing window enables the user to edit the calibration coefficients for every sub channel on the Slam Stick data logger 7 The coefficients window should display the coef...

Страница 33: ...ynomial coefficients in the same manner as the X and Y coefficients and click Apply to apply the correct coefficients This same process can be repeated for the piezoelectric accelerometer but this acc...

Страница 34: ...pdate the driver It may take as much as one minute but the LED will then turn red and a Silicon Labs CDC Serial Port should show up in your control panel Unplug the Slam Stick once this occurs Once th...

Страница 35: ...and vibration measurements for initial testing Film Television Stunt work for science programs All Slam Sticks are rated to operate between 20 C and 60 C Operation outside of this temperature range is...

Страница 36: ...hese two methods of mounting are highly recommended and encouraged other mounting options such as magnetic mounting are also acceptable An incorrect mounting configuration could result in erroneous da...

Страница 37: ...actice to use a washer If you have a plastic Slam Stick X or Slam Stick C the use of a washer is very important especially if you are using metal bolts The recommended mounting torque is 70 in oz for...

Страница 38: ...t these coefficients for already recorded data Mide will also update the firmware on the device if there is an update available to reflect any improvements or bug fixes The calibrated device will be s...

Страница 39: ...ow it to wait for the trigger or hold down button to skip to recording Rapid blinking Green Device is recording battery level is above 75 Allow it to continue recording or press button to end recordin...

Страница 40: ...to download online here There are several how to videos that quickly go through some data viewing and analysis functionality of the software on the website Figure 14 Video 2 of Mid s how to video seri...

Страница 41: ...OTE The software will be faster if loading files copied onto the computer and not opened directly from the device The viewing window and File drop down menu is shown in Figure 16 Figure 16 Opening an...

Страница 42: ...s were less dense and or less severe than in actuality The Buffer Maximum and Minimum display can help visualize how the maximum and minimum acceleration levels are changing with time This can be espe...

Страница 43: ...t are for the amplitude range The center button zooms to fit for that axis and there are keyboard shortcuts for this as well Alt 0 will zoom to fit in the y axis CTRL 0 will zoom to fit in the x axis...

Страница 44: ...f view options are accessible via the View menu Preferences There are a number of user preferences which can be set in the Preferences window accessible under the Edit menu Many of these options are s...

Страница 45: ...more information Drawing Options The user can opt between having one plot tab per sensor per channel or per sensor type There are also some numerical variables for controlling how many samples are pl...

Страница 46: ...seful tab in this window is the channel info tab which has max min values for each channel and when they occurred The sample rate is also shown along with the sensor part number Figure 20 The recordin...

Страница 47: ...the signal The user can either remove the mean across the entire recording or remove a rolling mean the time interval is editable under the preferences tab or disable mean removal to see the raw signa...

Страница 48: ...r degree C Disabling this temperature dependent calibration is useful for reducing computation time especially when your recording is within a couple degrees of room temperature Display Warnings Data...

Страница 49: ...disabled to expedite analysis time Figure 23 The analysis window is very similar for rendering plots FFTs PSDs and spectrograms this is accessible under the File Render drop down menu Render Plots Thi...

Страница 50: ...waves and we use Fourier analysis or spectrum analysis to deconstruct a signal into its individual sine wave components The result is acceleration vibration amplitude as a function of frequency which...

Страница 51: ...s is that this amplitude value is then normalized to the frequency bin width to get units of g2 Hz By normalizing the result we reduce the dependency on the number of samples in the original acquisiti...

Страница 52: ...window the FFT is applied to the result and the PSD is taken A windowed PSD uses less computational power and memory and results in cleaner looking data because it is an average of data sets Note the...

Страница 53: ...rmat One of the how to videos quickly gores through some examples of exporting data with the software on the website Figure 28 Video 4 of Mid s how to video series goes through some examples on export...

Страница 54: ...selecting what file type the range if you d like column headers and the time format you wish When exporting to MATLAB the array will likely be the transpose of what you may be used to this is done for...

Страница 55: ...ny recording files Figure 31 The tools allow the user to perform batch exports conversions and file splitting The dialog for all of these tools are very similar as shown in Figure 32 Each require the...

Страница 56: ...ty raw2mat discussed in the Command Line Executables for Converting IDE Files section The user can perform batch exports of IDE files to MAT but the data will be in its raw 16 bit integer form Support...

Страница 57: ...costs over 2 000 for a commercial license only 150 for a home license If cost is a concern Python offers an excellent alternative because it is open source and free Their SciPy library offers a colle...

Страница 58: ...e utility completed the conversion of a recording file with nearly 4 million data points in about 10 seconds When loaded into MATLAB you ll see a large number of variables as shown below These arrays...

Страница 59: ...or data This command line utility has the same functionality built into the Slam Stick Lab under the tools menu Calibrated Data Export Utility This utility is not as time efficient as the raw2mat util...

Страница 60: ...rmation shown in the legend in the Slam Stick Lab main plot window It took 5 times longer 50 seconds to generate the calibrated CSV files of that vibration recording Three separate files are generated...

Страница 61: ...storage capacity of the Slam Stick products can generate very large IDE files potentially large enough to adversely affect the performance of Slam Stick Lab and even MATLAB and Python scripts Mid deve...

Страница 62: ...is shifted back overlapping the end of the previous file The time of the first accelerometer sample will be correct relative to the last accelerometer sample MATLAB Example Functions Mid provides som...

Страница 63: ...er through how all this is being done SSL_Converter This function uses the command line utility raw2mat to take an IDE file and convert it to MAT then calls the SSL_Analyzer function to apply calibrat...

Страница 64: ...s incredibly useful and he has a lot of great content on his blog detailing how to use the package Some of the functions embedded in this free package are only available to paying customers of MATLAB...

Страница 65: ...User Manual Slam Stick User Manual Version No 2 0 65...

Страница 66: ...Mid has had good luck with the process outlined in this online video Load_Plot_RMS_FFT_Spectrogram Mid provides an example Python script named Load_Plot_RMS_FFT_Spectrogram py that can be run from the...

Страница 67: ...l Version No 2 0 67 Figure 33 An example python script to load a CSV plot data compute moving RMS compute a FFT and compute a spectrogram The script can be run from the command line as shown and it wi...

Страница 68: ...User Manual Slam Stick User Manual Version No 2 0 68 The plots generated from this script analyzing the car engine data are shown below...

Страница 69: ...wnloaded from his blog It reads in a CSV file that has just two columns time and the data of interest Use that example py python script to see how to take data from a CSV file with multiple sensor sub...

Страница 70: ...strate how an IDE file can be converted to XML for the more advanced user looking to directly load data in a recording file The scripts are well documented but contact Mid with any specific questions...

Страница 71: ...le is Corrupted or Does Not Display Data 0 KB file The Slam Stick may have suddenly ran out of power and therefore did not close out the file properly This can happen rarely due to battery dips but ca...

Страница 72: ...vice information Product Support For technical support repair and returns the best way to contact Mid is through our web contact page http www mide com pages contact and select Data Logging as the are...

Страница 73: ...and durability are paramount and semi permanent mounting for instance bolting or epoxy adhesive is available What is the MEMS DC accelerometer option and why do I need it The MEMS accelerometer or DC...

Страница 74: ...chargers that have a smart auto shut off feature once the device is charged You ll need a product that doesn t have this feature such as the V15 4 000 battery or V44 12 000 mAh battery from Voltaic W...

Страница 75: ...s own natural frequency during these events Some customers use mounting fixtures which aren t stiff enough which can amplify shock vibration events Lastly you can continue using the unit you have in t...

Страница 76: ...quick overview of different accelerometer types and provides some more information on integrating acceleration data What are the noise characteristics of the embedded accelerometers See the Component...

Страница 77: ...ery component axis that is a result of the on board programmable gate array doing the temperature compensation It s a characteristic of this accelerometer Figure 36 The piezoelectric accelerometer has...

Страница 78: ...f our technology and know how is centered on data acquisition and analysis So in the future our product line will evolve to accommodate external sensors and more embedded sensor options We also recogn...

Страница 79: ...selection to analysis Many links are provided to supplemental information and suppliers to keep the eBook as quick hitting and brief as possible Links are not limited to Mid s partners links to Slam...

Отзывы: