EM TEST
AN 200
Bedienungsanleitung
V 2.00
12 / 16
5.5.
Measuring results
5.5.1. Voltage drop in dependence of the test load
DUT current
voltage drop ( reference value )
Tolerance
1
A
1.8
V
±
30%
10
A
3.0
V
±
30%
23
A
3.5
V
±
30%
45
A
(only AN200B 60V50A)
4.5
V
±
30%
Î
manufacturer values
5.5.2. Measuring results with calibration according to standards
All measuring values indicated as well as the curve illustrated have been recorded with a serial model.
Reference values measured with AN 200
Input Voltage:
17.75 V
Output Voltage:
13.50 V
Upeak:
-
401 V
tf (90% - 10%):
323 ns
Figure 12 : Pulse parameters
Impedance setting 40
Ω
Va
output voltage
V
peak
tf
90% - 10%
fall time
Mit Normlast
-13.5V
±
10%
400V
±
30%
300ns
±
20%
Mit Normlast
( nur
AN200B 60V50A
)
-27V
±
10%
600V
±
30%
300ns
±
20%
Example for the switch off impulse
Figure 13 : Test load at 13.5V
Figure 14 : Test load at 13.5V 27V ( AN 200B )