Precision Ultrasound Balance
11
Controls and Markings
Refer to figure 1 - Assembled Balance.
Display:-
The liquid crystal display gives a direct reading of
ultrasonic power in Watts.
On/Off Button:- To turn the balance on, press the On/Off button.
Pressing either of the Tare buttons will zero the
display
To turn the balance off, press and hold down the
On/Off button.
Enter Button
When first turned on, the balance will default to show
grams. The Enter button will switch the display to the
customised output (Watts).
Select Button
This button MUST ONLY be used to toggle
between grams and watts AFTER the Enter
button has been pressed. Incorrect use will allow
access to and possible corruption of conversion
factors.
The EMS serial number of the balance and month / year of manufacture
is shown on the label attached at the side of the balance.