8
DirectShow
If you change the link ordering after installation you must run dplinks.exe. This program will
update the existing input names used by the windows DirectShow interface.
To run the dplinks program open the
Run
by clicking on
Start/Run
and type dplinks and then press
Enter.
The program will run, however no notifications are displayed.
Firmware Upgrades
The
X
TREME
RGB-SDI2
cards allow
firmware
upgrade to be completed on site rather than returning
the card to EMS. Whenever a
firmware
upgrade is performed,
LK1 MUST BE FITTED on the
X
TREME
RGB-SDI2.
To perform the upgrade, follow the step-by-step instructions provided by the upgrade application.
In the unlikely event that something goes wrong during the upgrade process (e.g. System power
outage) it is possible to revert to the factory settings by powering down the system, temporarily
removing LK1 then powering up the system with the link removed. Once the system has
rebooted, replace the LK1 link (whilst the system is powered up) and restart the
firmware
upgrade process.
It should be noted that the latest driver installation program includes an automatic
firmware
update, if required. Therefore, prior to installing the application and driver, ensure that LK1 is
fitted.
Software Installation
The Xtreme software (driver and application) is installed at
www.ems-imaging.com/downloads
Install Software…
and follow the installation wizard instructions as prompted.
Regular software updates are available from our website:
www.ems-imaging.com/downloads