Phoenix User’s Manual

Low Voltage Indication

• Indicator Threshold: 2.3 Volts (typical)
• Shutdown Voltage: 1.95 Volts (typical)
• These voltages may be tested under NO load condition.

Fuse Characteristics

6V, 1.5 A, Resettable, breaking capacity 3A

Physical Characteristics

• Size (without belt clip): 5.4” x 2.6” x 1.2”
• Weight (with batteries and belt clip): ~ 6 ounces
• Operational Temperature: 10°C to 40°C
• Humidity (maximum): 75% RH

EN 60601-1 Classification

• Type BF Applied Part
• Internally powered only
• Protection against ingress of particles and liquids:  
• Continuous operation
• Not suitable for use in the presence of a flammable anesthetic mixture with air or with 

oxygen or nitrous oxide. 

Содержание phoenix

Страница 1: ...Empi CONDUCTIVE GARMENT USER S MANUAL Read carefully before using this Garment Visit www djoglobal com...

Страница 2: ...ian or other practitioner licensed to use the device Intended Use The Empi Phoenix Conductive Garment is intended for use with the Empi Phoenix device which is indicated for retarding or preventing di...

Страница 3: ...ly fastened to prevent inadvertent disconnection Electrodes will eventually wear out Check accessories regularly for signs of wear and replace as needed Skin reactions On rare occasions therapy can re...

Страница 4: ...should only be placed within the corresponding larger grey white dashed outline IMPORTANT Make sure to place ALL THREE electrodes on the garment before using The electrodes must completely cover each...

Страница 5: ...ust remove the garment completely and reposition 8 Connect the black plastic leadwire connector to the garment by sliding it onto the port as shown You should hear an audible click 9 Attach the Empi P...

Страница 6: ...ctrode Use section for Phoenix Garment Electrodes maintenance suggestions When electrodes are no longer useful dispose of them in a receptacle out of the reach of children or pets If needed you may wa...

Страница 7: ...ally approved DJO Global service center This warranty is in lieu of any or all other warranties expressed or implied No person is authorized to bind Empi to any representative of warranty other than t...

Страница 8: ...Manufactured in Mexico for Empi Inc 205 Hwy 22 East Clear Lake SD 57226 USA 1 651 415 9000 360412 Rev C 2014 Empi EU Authorised Representative MDSS GmbH Schiffgraben 41 30175 Hannover Germany...

Страница 9: ...Empi COMPLETE ELECTROTHERAPY SYSTEM USER S MANUAL Read this manual carefully before operating the Phoenix Visit us at www djoglobal com...

Страница 10: ...trode Positions 15 6 4 1 Use of the Empi Phoenix Thigh Garment 15 6 5 Stimulation Positions 16 6 6 Adjusting Stimulation Energies 16 7 Operating Instructions 17 7 1 Description of the Device 17 7 2 Ki...

Страница 11: ...ation 26 9 3 Repair 27 9 4 Operating Conditions 27 9 5 Transportation and Storage Conditions 27 9 6 Expected Life and Disposal 27 10 Ordering Information 28 11 Clinician Only Section 28 11 1 Complianc...

Страница 12: ...pecific muscles or muscle groups to help treat disuse atrophy and re educate muscles The pre set programs are designed to provide therapeutic benefit while minimizing complexity for the patient and cl...

Страница 13: ...te pain Relief of pain associated with arthritis As a pulsed current device indications are for the following conditions Reduction of edema under negative electrode Reduction of muscle spasm Influenci...

Страница 14: ...of 12 5 mm 0 5 in diameter and greater Protected against vertically falling water drops when enclosure tilted up to 15 Minimum and maximum temperature indications to respect Prescription only USA C E...

Страница 15: ...or blood pressure Stimulation across chest Do not apply stimulation across the patient s chest because the introduction of electrical current into the chest may cause rhythm disturbances to the patien...

Страница 16: ...ce only with the leads electrodes and accessories recommended by Empi Use of other accessories may adversely affect the performance of the device or may result in stronger electromagnetic emissions or...

Страница 17: ...and the battery compartment Garment Do not use the Empi Phoenix Thigh Garment in proximity of fire or excessive heat sources due to the risk of fire Make sure that the electrodes cover the metal conn...

Страница 18: ...figurations will deliver a charge of up to 20 microcoulombs C or greater per pulse and may be sufficient to cause electrocution Electrical current of this magnitude must not flow through the thorax be...

Страница 19: ...be achieved using a pulsed direct current The Empi Phoenix program P4 is a pulsed direct current program During voluntary activity the brain sends a command to the nerve fibers in the form of an elect...

Страница 20: ...of time This program is recommended for use before and after the surgery as prescribed by your medical professionals P1 begins with a two minute Warm Up phase which will count down on the screen The W...

Страница 21: ...the Work phase begins the intensity will automatically decrease by half You can then adjust the intensity to provide a comfortable but strong muscle contraction The program finishes with a Cool Down...

Страница 22: ...ction and increased circulation For acute Edema less than 48 hours post injury or post op place the positive electrodes over the treatment site and attach the negative leadwire pins to the dispersive...

Страница 23: ...ensities more than 2mA cm2 Precaution Do not apply stimulation in the vicinity of metal Remove jewelry body piercings buckles or any other removable metallic product or device in the area of stimulati...

Страница 24: ...ion Never carry out an initial stimulation session on a person who is standing The first five minutes of stimulation must always be performed on a person who is sitting or lying down In rare instances...

Страница 25: ...is an optional accessory J Belt Clip removable K Battery door to access the battery compartment A label describing the programs is affixed to the battery door 7 Operating Instructions You are strongl...

Страница 26: ...for Use 360413 1 Phoenix Instruction for use Quick Start Guide 802425 1 Quick start guide for the patient DVD 802427 1 DVD OPTIONAL PHOENIX THIGH GARMENT Phoenix thigh garment 235684 1 Garment Garment...

Страница 27: ...the device 2 Remove the battery door by sliding it downward along the device 3 Place the batteries as indicated on the bottom of the battery compartment Note the polarity and the battery type Use only...

Страница 28: ...electrodes After cleaning thoroughly rinse with water and dry the skin carefully Dry electrodes with poor adhesion can be reconditioned See electrode packaging for complete instructions If you encount...

Страница 29: ...therapy timer is activated the device switches off automatically at the end of the programmed interval The remaining therapy time is always indicated on the display The program can only be changed wh...

Страница 30: ...h off the unit by pressing the ON OFF button a second time 7 4 5 End of treatment For all treatments except P3 TENS the timer will automatically finish the treatment When the treatment is finished int...

Страница 31: ...art of the symbol will blink TENS P3 and Edema P4 treatments do not have a rest mode 7 5 4 Timer indication The timer indicates the remaining time of the active sequence For the TENS program P3 the ti...

Страница 32: ...mulation replace the lead wires and or electrodes Display does not come on Battery A Try fresh batteries B Ensure batteries are inserted correctly See instructions for proper placement Battery contact...

Страница 33: ...Improper placement A Replace B Reposition electrode Electrodes must be a minimum of 2 inches apart Leadwires worn or damaged Replace Component failure Try each leadwire independently in each channel I...

Страница 34: ...act Empi at the numbers below if you need assistance setting up using or maintaining the Phoenix System or to report any unexpected operation or events For clinical questions contact the Professional...

Страница 35: ...relative humidity between 10 and 90 9 6 Expected Life and Disposal The Empi Phoenix device is expected to provide at least seven years of normal use If you are renting the Empi Phoenix device please...

Страница 36: ...Right Decrease Intensity buttons while you are in the Program Selection mode In the compliance screen you will see the total time spent in each of the programs by switching between the different prog...

Страница 37: ...ntinuously and simultaneously on the Program Increase Program Decrease and Right Decrease buttons 3 The Program Lock symbol will appear The lock setting is kept in memory when you turn off the device...

Страница 38: ...written notice of the defect setting forth the problem in reasonable detail prior to expiration of the Warranty Period and within 30 days of discovery of the defect Upon Empi Inc s written request and...

Страница 39: ...Empi Inc has any authority to bind Empi Inc to any affirmation representation or warranty except as stated in this written warranty policy D Empi Inc shall not be liable to any person for any direct i...

Страница 40: ...A independent of the set intensity as long as the set intensity is above 4 mA Maximum charge per pulse 7 2 C for the positive pulse and 6 C for the negative pulse 13 Technical Specifications Standard...

Страница 41: ...ty 3A Physical Characteristics Size without belt clip 5 4 x 2 6 x 1 2 Weight with batteries and belt clip 6 ounces Operational Temperature 10 C to 40 C Humidity maximum 75 RH EN 60601 1 Classification...

Страница 42: ...properly in the chosen configuration Meeting the emissions levels shown in the first table is considered to be essential performance of the Empi Phoenix device GUIDANCE AND MANUFACTURER S DECLARATION...

Страница 43: ...l commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode Not Applicable Battery powered Mains power quality should be that of a typical commercial or hospital e...

Страница 44: ...RF transmitters as determined by an electromagnetic survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the fol...

Страница 45: ...on distances according to frequency of the transmitter m From 150 kHz to 80 MHz d 1 2 P From 80 kHz to 800 MHz d 1 2 P From 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 20 1 2...

Страница 46: ...thigh 5 To use on a portion of the body other than the knee quadriceps consult a clinician TENS Program P3 High Frequency TENS 1 Place the Channel 1 black lead over the upper inside portion of your k...

Страница 47: ...ton 5 Adjust intensity to desired level by pressing the and buttons for channels 1 and 2 You must select your program before turning up the intensity Once the intensity is increased the program will b...

Страница 48: ...Empi Inc 205 Hwy 22 East Clear Lake SD 57226 USA 651 415 9000 800 328 2536 360413 Rev A 2013 Empi Inc 04 13 199696 Phoenix Kit...

Отзывы: