Phoenix User’s Manual
32
P1 Endurance Program
Maximum Intensity:
520 Energy levels,
corresponding to 100 mA intensity and 300 µs
pulse width
Zero net DC component
Maximum charge per pulse:
2 x 30 µC
P2 Strength Program
Maximum Intensity:
520 Energy levels,
corresponding to 100 mA intensity and 300 µs
pulse width
Zero net DC component
Maximum charge per pulse:
2 x 30 µC
P3 Modulated TENS Program
Maximum Intensity:
100 mA
Zero net DC component
Maximum charge per pulse:
2 x 8 µC
P4 Edema Program
Maximum Intensity:
100 mA
DC component:
266 µA, independent of the set
intensity as long as the set intensity is above 4 mA.
Maximum charge per pulse:
7.2 µC for the
positive pulse and 6 µC for the negative pulse
13. Technical Specifications
Standard Measurement Conditions
• Temperature - 25°C +/-5°C
• Load Range - 500 Ohms - 1 kOhms
• Power Supply - 3.0V DC +/-10%
Output Waveforms
The following are theoretical standard measurement output current across purely resistive
loads at maximum intensity setting. Pulse Width and current measured as shown across
1kOhm loads. These measurements are also valid on a 500 Ohm load, as the Phoenix is a
current controlled device. Any load between 500 Ohms and 1 kOhm will not affect the output
measurements. Your output may vary depending on parameter settings.