Testing High Circuit

 – If the tester’s high range is 

35Meg. Set the knob on model 50421 to “35M PASS”, 
touch the TEST contact on the tester and you should 
get a green LED for the left and right feet. Set the 
knob on model 50421 to “35M HIGH”, touch the TEST 
contact on the tester and you should get a yellow LED 
for both the left and right feet. If the limit is set to 1Gig 
on the tester, test at 1Gig on model 50421, same for 
10Meg and 100Meg. Disregard the test result for the 
wrist strap, if the wrist strap circuit is on.
 

Testing Wrist Strap Circuit, refer to fi gure 2

To complete 
the wrist strap 
test, you will 
need to test 
the low and 
high limits. 
Refer to the 
dipswitch 
setting on 
the left side 
of the testers 
for wrist strap 
test ranges. 
Manufacturer’s suggested default test range is 750K 
low and 10Meg high for the US and 750K to 35Meg for 
Europe.

Do not power down the tester. Using the 2 banana 
leads included with model 50421. Attach 1 of the 
banana leads to the right side banana jack on model 
50421 and connect the other end to the ground sym-
bol jack on the tester. Connect the last banana jack to 
the left banana jack of model 50421 and the other end 
to the “SINGLE – WIRE” banana jack. Note be sure 
FOOTPLATE lead is removed.

Testing Low Circuit

 - If the tester’s low range is set to 

750K. Set the knob on model 50421 to the “750K LOW” 
position. Touch the TEST contact down on the tester 
and you should get a red LED for the Wrist strap. Dis-
regard the test result for the footwear. Set the knob on 
model 50421 to “750K Pass”, touch the TEST contact 
on the tester and you should get a green LED for the 
wrist strap.

Testing High Circuit – 

If the tester’s high range is 

10Meg. Set the knob on model 50421 to “10M PASS,” 
touch the TEST contact on the tester and you should 
get a green LED for the wrist strap. Set the knob on 
model 50421 to “10M HIGH”, touch the TEST contact 
on the tester and you should get a yellow LED for the 
wrist strap. If the limit is set to 35Meg on the tester, test 
at 35Meg on model 50421. Disregard the test result for 
the footwear.

SPECIFICATIONS AND DIMENSIONS 

Charge Detection

When the operator is connected to the remote and 
accumulates a voltage greater than ± 1.25 volts, the 
CHARGE LED (red) will display in its respective OP-
ERATOR section (for charge detection data out, see 
the Data Output section on Page 8).  To disable charge 
detection on the display, shunt jumper JP6 inside the 
unit (See Page 7).  If a charge is detected while using 
this confi guration, the display will remain in its previous 
state until the charge is no longer detected.

Testers limit comparator will calibrate:

50404

 - Dual Independent Footwear Tester,

 Foot 

Plate,120V 

50412

 - Dual Independent Footwear Tester, No Adapter,  

 CE, 

Bin

50407

 - Dual Independent Footwear and Wrist Strap  

 

Tester, USA Standard

50413

 - Dual Independent Footwear and Wrist Strap  

 

Tester, Euro Standard

50562

 - Dual Independent Footwear and Wrist Strap  

 

Tester, Dual Foot Plate, Euro Standard

50430

 - SmartLog X3, w/Dual Foot Plate, Keypad &  

 

Barcode Reader, Software, 120V

50431

 - SmartLog X3, w/Dual Foot Plate, Keypad &  

 

Barcode Reader, No Software, 120V

50432

 - SmartLog X3 w/Dual Foot Plate, Keypad, &  

 

Barcode Reader, Euro Specs

50434

 - SmartLog X3 w/Dual Foot Plate, Laser 

 

Scanner, Software, Euro Specs, CE Approved