Calibration

The Surface Resistance Test Kit is calibrated to NIST 
traceable standards.  Most users require calibration 
annually. Please call our Customer Service Department at 

909-664-9980 (Chino, CA) or 781-821-8370 (Canton, MA) 

for details. In-house calibration can be performed by using 

1% resistors in each of the meter ranges. Simply attach the 

resistors to the cords using grounding clips and recording 
the meter display. Keep the cords separated.   Should 
adjustment be necessary, it is recommended that the unit 
be returned to the factory as access to internal adjustments 
requires special equipment and test cables.

CALIBRATION EQUIPMENT:

•  Digital Multimeter - accurate to 1.25% @ 10VDC and 

100VDC

•  Fixed Decade Box value 10E3 - 10E12 - accurate to 

±2.5%, except at 10E11 and 10E12 (±5%)

•  Thermometer - accurate to ±1°F

•  Humidity meter - accurate to 2%

•  Test leads

•  99% Isopropyl alcohol and cleaning wipes

SET UP:
A. Test Area

 - Area needs to be free of any high voltage 

transformer or power supply. Not to be performed under 
any type of fluorescent lighting or high power lighting.

B. Worksurface

 - needs to be covered with conductive mat 

at 1.0 x 10E3 or less, connected to equipment ground.

C. Technician

 - needs to be grounded with zero ohm 

resistor to equipment ground.

D. Decade Box 

- needs to be grounded to equipment 

ground.

Accuracy is measured after normalizing the instrument at 

70-80 °F and 30-50% RH for a minimum of 4 hours.

NORMALIZATION OF SURFACE RESISTANCE TEST KIT

Temperature inside testing area needs to be 75°F @ 

±6.6% 40% to 60% RH. Testers need to stay at constant 

temperature 75°F @ ±6.6% for about 2 hours for proper 

reading. Meter cannot be inside objects, enclosed boxes, 

containers or cases unit is supplied with (temperature inside 
case will differ from outside temperature, cases will act like 
an insulator to the Meter), 

Meter will have to be stationary 

in testing area for about 2 hours with no significant 
temperature changes.

CALIBRATION TEST PROCEDURE

•  When testing use only the leads supplied with the meter. 

DO NOT TEST WITH RESISITIVITY ATTACHMENT, 

Resistivity Attachment will give a much higher reading 
and is only designed for spot checking.

•  With 10V/AUTO switch down, press TEST button - 

voltage between the two leads should be 10V ±5%

•  With 10V/AUTO switch up, press TEST button - voltage 

between the two leads should be 100V ±5%

•  Using the cleaning wipes and 99% isopropyl, clean 

around banana jack and mono jack where leads connects 
to, oil from human fingers can alter accuracy.

•  Make sure 10V/AUTO switch is set to AUTO (switch up). 

Testing each decade starting from 1.0 x 10E12 and down, 

do not start from 1.0 x 10E3.

Temperature  

Fahrenheit  

= 75°F ±6.6%

 

 

Celsius                = 23.8°C ±10.2%

Relativity Humidity ±10 Digits 

A. 1.0 x 10E12

 

+ 20%   LED = 12 Yellow Mantissa 1.20

 

0% 

LED = 12 Yellow Mantissa 1.00

 

- 20%   LED = 11 Yellow  Mantissa 8.00

B. 1.0 x 10E11

 

+20%  LED = 11 Yellow Mantissa 1.20

 

0% 

LED = 11 Yellow Mantissa 1.00

 

-20%  LED = 10 Green  Mantissa 8.00

C. 1.0 x 10E10

 

+10%  LED = 10 Green  Mantissa 1.10

 

0% 

LED = 10 Green  Mantissa 1.00

 

-10%  LED =   9 Green  Mantissa 9.00

D. 1.0 x 10E9

 

+10%  LED =   9 Green  Mantissa 1.10

 

0% 

LED =   9 Green  Mantissa 1.00

 

-10%  LED =   8 Blue  Mantissa 9.00

E. 1.0 x 10E8

 

+10%  LED =   8 Blue  Mantissa 1.10

 

0% 

LED =   8 Blue  Mantissa 1.00

 

-10%  LED =   7 Blue  Mantissa 9.00

F. 1.0 x 10E7

 

+10%  LED =   7 Blue  Mantissa 1.10

 

0% 

LED =   7 Blue  Mantissa 1.00

 

-10%  LED =   6 Blue  Mantissa 9.00

G. 1.0 x 10E6

 

+10%  LED =   6 Blue  Mantissa 1.10

 

0% 

LED =   6 Blue  Mantissa 1.00

 

-10%  LED =   5 Green  Mantissa 9.00

H. 1.0 x 10E5

 

+10%  LED =   5 Green  Mantissa 1.10

 

0% 

LED =   5 Green  Mantissa 1.00

 

-10%  LED =   4 Green  Mantissa 9.00

I. 1.0 x 10E4

 

+10%  LED =   4 Green  Mantissa 1.10

 

0% 

LED =   4 Green  Mantissa 1.00

 

-10%  LED =   3 Red  Mantissa 9.00

J. 1.0 x 10E3

 

+10%  LED =   3 Red  Mantissa 1.10

 

0% 

LED =   3 Red  Mantissa 1.00

 

-10%  LED =   <3 Red  Mantissa <0.90

EMIT

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Fax (909) 627-7449 • Website: 

DescoEMIT.com

Содержание 50557

Страница 1: ...ting Resistive Measurement Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versati...

Страница 2: ...rees Celsius tolerance 3 C typical Humidity as percentage from 5 95 tolerance 10 typical Surface resistance mantissa with exponent displayed via illuminated LED measurement in ohms Measurement being d...

Страница 3: ...S4 1 Worksurfaces The test specimens and electrodes shall be cleaned twice with a minimum 70 isopropanol water solution using a clean low linting cloth each time Note then conditioned for 72 hours min...

Страница 4: ...ensitive devices shall also be removed The black sensing test lead with shielded plug should go to equipment ground or a groundable point Use one Electrode on other test lead and place Electrode the f...

Страница 5: ...hms RTT Surface Resistivity should measure 10E8 ohms square Ref ANSI ESD S11 11 section 12 0 Conversion to Resistivity states When it is appropriate to convert a resistance obtained by this test metho...

Страница 6: ...ses will act like an insulator to the Meter Meter will have to be stationary in testing area for about 2 hours with no significant temperature changes CALIBRATION TEST PROCEDURE When testing use only...

Страница 7: ...f your original packing slip invoice or other proof of date of purchase Any unit under warranty should be shipped prepaid to the EMIT factory Warranty replacements will take approximately two weeks If...