background image

TB-6520

 Page 3 of 7

Surface resistance ohm values are expressed with a 

mantissa and exponent or power of the number. For 

example, if “8” is illuminated by its LED and the LCD 

displays a mantissa of “7.14”, the measurement is 7.14 x 

10E8 ohms or 714,000,000 ohms

If surface is over 10E12 ohms, the following will appear:

•  “1____” on the LCD display indicate Overrange or that the 

reading exceeds the display capabilities

C. Test Button:

 This red button activates electrical power 

to the Meter.  The exponent or power of the number is 

displayed immediately with LED illuminating number.  If “8” 
is illuminated, the measurement is in the 10

8

 or 10E8 ohm 

range.  To make a measurement in accordance with 

ANSI/ESD S4.1, the button is to be depressed for 10-20 

seconds for the unit to make numerous readings and 
calculations.  (Testing in accordance with ANSI/ESD S4.1 

requires 15 seconds of electrification; in contrast, most 

analog type meters display measurements instantaneously.) 
The LCD will display temperature (F), then temperature 
(C), then relative humidity, and then the surface resistance 

mantissa.  For example, if the LCD displays a mantissa of 

“7.14”, and “8” is illuminates by its LED, the measurement is 

7.14 x 10E8

 

ohms or 714,000,000 ohms. 

During the entire period, surface resistance exponent will be 

displayed by LED illuminating number.

D. Override Test Voltage Button:

 When in the “up” 

position, during resistance portion of test, Meter will 
automatically switch to the correct voltage for the resistance 
range.  LED will illuminate noting selected voltage.  

Material 10E5

 

ohms or less should be measured at 10 

volts.  Material 10E6 ohms or greater should be tested 

at 100 volts.  The button is a switch, which if depressed 
will override automatic voltage selection and test will be 
performed at 10 volts regardless of resistance level.

E. Jacks:

 One end of Test Leads has 3.5mm plug (fits 

Meter left jack).  The 3.5mm plug is shielded (identified 

by insulated tip and barrel black lead) - see RTG test 

procedure: Per ANSI/ESD S4.1 “The sensing lead of the 
resistance meter shall be connected to groundable point.” 

One end of Test Lead has standard banana plug (fits 5# 

Electrode jack).

F. AA Battery Compartment.

G. Resistivity Parallel Electrodes Attachment. 

(optional Item number 

50560

)

Cleaning

Per ANSI/ESD S4.1 “Clean the electrodes with a minimum 

70% isopropanol-water solu tion.” Make sure conductive 

pads are dry prior to use. 

See specific product test standard for test lab specimen 

cleaning instructions.  Per ANSI/ESD S4.1 Worksurfaces 

“The test specimens and electrodes shall be cleaned twice 

with a minimum 70% isopropanol-water solution using a 

clean, low-linting cloth each time.” (Note: then conditioned 

for 72 hours, minimum).

For installed product periodic testing, do not clean surfaces.  
However, if any measurements lie outside acceptable 

range, then clean the surface and re-test (Note: for 

worksurfaces, use an ESD cleaner not containing silicone). 

Be sure the surface is dry before testing.

Periodic maintenance

 - The area surrounding the cable 

jacks at the top end of the meter should be wiped with a 
clean cloth moistened with alcohol to remove skin oils that 
will accumulate and affect the accuracy at high resistances.  

The frequency of cleaning will depend on usage;  once a 

month would be a good starting point.  Other items that 

should also be cleaned in this fashion are the cable jackets 
and the resistivity attachment, if included.

Power Requirements

The Meter is powered by two replaceable alkaline AA 
batteries.

Battery Replacement

Depress buttons on both sides of bottom end cap of meter 
housing. Remove cap, press on right edge of protruding 
battery door, unlatching it, and swing door open. Carefully 
replace batteries with alkaline Type AA. Polarity must be 
correct or damage may occur. Close battery door and 

re-install bottom end cap on meter housing.

Test Procedure
General Guidelines:

•  Use both 5 Pound Electrodes for RTT

•  Use one 5 Pound Electrode and black sensing lead to 

equipment ground to groundable point for RTG 

•  Use optional Resistivity Attachment (removing leads & 

electrodes) for Resistivity measurements

•  Ensure that item being measured is electrically isolated 

(i.e. placed on an insulative surface) as Meter will 
measure lowest resistance path

•  Ensure that test leads are separated or Meter may 

measure lower resistance path

•  When using 5 Pound Electrodes:

  •  Place no closer than 2" from edge of surface being  

  measured.

  •  Place no closer than 3" to any groundable point

  •  Place 5 Pound Electrodes about 10" apart for RTT of 

    worksurfaces and about 3' apart for RTT of flooring.

  •  Preferred placements include: most commonly used  

  surface portion, most worn, center, and furthest from  
  groundable point.

•  If surface has sections (like floor tiles or garment panels), 

for RTT place a 5 pound electrodes on different sections

•  Clean surface for Test Lab, but do not initially clean 

surface for installed products (if fails, clean and retest).

•  Only use an approved ESD cleaner and allow to dry prior 

to testing. 

•  Lay Meter on surface; holding in hand can cause 

measurent distortions.

© 2010 DESCO INDUSTRIES INC.

Employee Owned

EMIT

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Fax (909) 627-7449 • Website: 

DescoEMIT.com

Содержание 50557

Страница 1: ...ting Resistive Measurement Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versati...

Страница 2: ...rees Celsius tolerance 3 C typical Humidity as percentage from 5 95 tolerance 10 typical Surface resistance mantissa with exponent displayed via illuminated LED measurement in ohms Measurement being d...

Страница 3: ...S4 1 Worksurfaces The test specimens and electrodes shall be cleaned twice with a minimum 70 isopropanol water solution using a clean low linting cloth each time Note then conditioned for 72 hours min...

Страница 4: ...ensitive devices shall also be removed The black sensing test lead with shielded plug should go to equipment ground or a groundable point Use one Electrode on other test lead and place Electrode the f...

Страница 5: ...hms RTT Surface Resistivity should measure 10E8 ohms square Ref ANSI ESD S11 11 section 12 0 Conversion to Resistivity states When it is appropriate to convert a resistance obtained by this test metho...

Страница 6: ...ses will act like an insulator to the Meter Meter will have to be stationary in testing area for about 2 hours with no significant temperature changes CALIBRATION TEST PROCEDURE When testing use only...

Страница 7: ...f your original packing slip invoice or other proof of date of purchase Any unit under warranty should be shipped prepaid to the EMIT factory Warranty replacements will take approximately two weeks If...

Отзывы: