ESD Control items should be tested:

A.  Prior to installation to qualify for listing in user’s ESD 

control plan approved ESD control items

B.  During initial installation

C.  For periodic checks of installed products as part of ESD 

control plan compliance verification plan

“A Compliance Verification Plan shall be established 

to ensure the Organization’s fulfillment of the technical 

requirements of the ESD Control Program Plan. Process 
monitoring (measurements) shall be conducted in 
accordance with a Compliance Verification Plan that 
identifies the technical requirements to be verified, the 
measurement limits and the frequency at which those 
verifications shall occur. The Compliance Verification Plan 
shall document the test methods and equipment used 
for process monitoring and measurements. If the test 

methods used by the Organization differ from any of the 

standards referenced in this document, then there must 
be a tailoring statement that is documented as part of 
the ESD Control Program Plan. Compliance verification 
records shall be established and maintained to provide 
evidence of conformity to the technical requirements.  The 
test equipment selected shall be capable of making the 
measurements defined in the Compliance Verification Plan.” 

(ANSI/ESD S20.20-2007 section 7.3)

Electrification Period

The Surface Resistance Test Kit provides the proper 

electrification period of 15 seconds per ANSI/ESD S4.1, 

after numerous readings and calculations are executed, 

then displays surface resistance mantissa measurement 
(Note: most analog type meters display measurements 
instantaneously).

When the Test Button is depressed, the liquid crystal 

display (LCD) will indicate:

•  Temperature in degrees Fahrenheit (tolerance ±5°F, 

typical)

•  Temperature in degrees Celsius  (tolerance ±3°C, typical)

•  Humidity as percentage (from 5% - 95% tolerance, ±10 

typical)

•  Surface resistance mantissa (with exponent displayed via 

illuminated LED, measurement in ohms)

Measurement being displayed is identified by the illuminated 

function LED. The surface resistance exponent (or power of 

number) is immediately illuminated and remains illuminated 
measuring the range the surface resistance in ohms.

Reference Literature

In addition to those noted above:

ANSI/ESD S20.20 - Development of ESD Control Program

ESD ADV1.0 - Glossary of Terms

ANSI/ESD S6.1 Grounding

These documents can be obtained directly from the 

ESD Association, 7902 Turin Rd., Suite 4, Rome, NY 

13440-2069, (315) 339-6937 or www.esda.org.

Other standards are available from the agencies who 

produce them. MIL-HDBK 263A, EIA-IS-5-A,  ASTM-F-150, 

EN 100015, and JESD625-A.  

Features

© 2010 DESCO INDUSTRIES INC.

Employee Owned

A. LED Displays:

 Surface resistance exponent is displayed 

via light emitting diodes illuminating 1/8” high number. 

There are 12 LEDs displaying surface resistance exponent 

measurement. They are color coded for quick checks:

 

Exponent 

Color

  <3, 3 

Red

  4, 5 

Green

  6, 7, 8 

Blue

  9, 10, 11 

Green

  12, >12 

Yellow

Five Function LED’s identify the measurement being 

displayed (see B below).
•  When battery voltage drops to approximately 2 volts, one 

of the Function LED’s to the right of the red pushbutton 
switch will begin to flash, indicating the need to replace 
batteries.

B. Liquid Crystal Display (LCD):

 Mantissa is displayed via 

9/16" high liquid crystal  display and provides easy to read 

resistance (or resistivity) measurements.

When the Test Button is depressed, the LCD will indicate:
•  Temperature in degrees Fahrenheit

•  Temperature in degrees Celsius 

•  Humidity as percentage 

•  Surface resistance mantissa (with exponent displayed via 

LED, measurement in ohms)

Figure 2. Features of the Surface Resistance Meter

EMIT

 - 3651 Walnut Avenue, Chino, CA 91710 • (909) 664-9980 • Fax (909) 627-7449 • Website: 

DescoEMIT.com

A

B

C

D

E

F

G

Содержание 50557

Страница 1: ...ting Resistive Measurement Floor Footwear ANSI ESD STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person Workstations ESD ADV53 1 ESD Protective Workstations Versati...

Страница 2: ...rees Celsius tolerance 3 C typical Humidity as percentage from 5 95 tolerance 10 typical Surface resistance mantissa with exponent displayed via illuminated LED measurement in ohms Measurement being d...

Страница 3: ...S4 1 Worksurfaces The test specimens and electrodes shall be cleaned twice with a minimum 70 isopropanol water solution using a clean low linting cloth each time Note then conditioned for 72 hours min...

Страница 4: ...ensitive devices shall also be removed The black sensing test lead with shielded plug should go to equipment ground or a groundable point Use one Electrode on other test lead and place Electrode the f...

Страница 5: ...hms RTT Surface Resistivity should measure 10E8 ohms square Ref ANSI ESD S11 11 section 12 0 Conversion to Resistivity states When it is appropriate to convert a resistance obtained by this test metho...