INSTALLING BATTERIES AND BATTERY CONSUMPTION 

Product operates with 2 pcs AA size 1.5 V batteries. Install and remove the batteries as follows: 
 

 
 
 
 
 
 
 
Estimated battery life is 6 months, when using high quality alkaline batteries, with 2800 mAh capacity (2pcs). Estimation is 
based on measured battery consumption in various conditions and then a calculation where device is on 50% of time (shut down 
50% of time) and of that 50% of time there is someone in bed 75% of that time, there are two notifications per day and sound 
notification is on 30 seconds each time.   
 
 

WHEN THE BATTERIES ARE GETTING LOW, THE RED LED BEGINS TO FLASH SLOWLY, A “BEEP” SOUND WILL BE HEARD 
AFTER EVERY 1,5 HOURS AND THE DRY-CONTACT OUTPUT GIVES LOW BATTERY NOTIFICATION AFTER EVERY 3 HOURS. 

 
 

9.

 

ABOUT THE OPTIONAL 5V AC ADAPTER 

For preparing the optional power supply for use, do the following: 

 
 
 
 
 

THE EMFIT EPILEPTIC TONIC-CLONIC SEIZURE ALARM IS DESIGNED AND TESTED TO BE USED ONLY WITH GLOBTEK 
INC. POWER SUPPLY MODEL: NO GTM41060-1505 AND P/N:WR9QA3000LCP-N-MNK. ANY OTHER TYPE OF AC ADAPTER 
MAY AFFECT THE PRODUCT’S SAFETY.  

 
The power supply is equipped with a blue indicator light; when the light is on the power supply is in use. If the light is off and 
the power supply is connected to the mains outlet, the power supply is probably damaged and should be replaced.  
 

WHEN THE AC ADAPTER IS CONNECTED TO THE POWER INPUT, THE BATTERIES ARE ELECTRICALLY DISCONNECTED.  
IN CASE OF NO MAINS POWER AND AC ADAPTER CORD IS CONNECTED TO X1, THE BATTERIES INSIDE WILL NOT 

OPERATE AS BACK UP POWER SUPPLY.  

 

 
 

Open the cover by lifting from 
one side 
 

Install 2 pcs good quality alkaline 
AA size 1,5 V alkaline batteries 
according the polarity drawings 
on the circuit board  

When removing old batteries, 
they are easiest to remove by 
lifting from + ends 
 

Remove the plastic cover  
 

Pick up suitable adapter from the 
included 4 pcs  
 

Install the adapter and make 
sure it stays in place  
 

Содержание D-1090-2G

Страница 1: ...L CARE OR THAT WOULD LIKELY LEAD TO A POTENTIALLY LIFE THREATENING SITUATION Applicable model numbers Control Unit D 1090 2G version t43 v 1 0 10 50Hz or newer Under Mattress Bed Sensor L 4060SL Date June 27th 2008 V2 2 Emfi Emfit and Emfit logo are either registered trademarks or trademarks of Emfit Ltd in EU USA Japan and or other countries Copyright 2002 2008 Emfit Ltd All rights reserved Paten...

Страница 2: ...LECTROMAGNETIC CONDITIONS 13 21 MANUFACTURER S CONTACT INFORMATION 15 DANGER NOTES THE FOLLOWING NOTES ARE PROVIDED BOTH FOR YOUR PERSONAL SAFETY AND TO PROTECT THE DESCRIBED PRODUCT OR ATTACHED DEVICES FROM DAMAGE SAFETY NOTES AND WARNINGS FOR THE PREVENTION OF DANGER TO THE LIVES AND HEALTH OF USERS OR MAINTENANCE PERSONNEL AND OR FOR THE AVOIDANCE OF DAMAGE TO PROPERTY ARE EMPHASISED IN THESE I...

Страница 3: ...tems for transferring the notification via dry contact opto coupler A reset switch for turning off the notification sound is to the right of the sensor input marked as SW1 Inside the device there is a place for 2 pcs AA size 1 5 V batteries and mini switches DIP and rotary for program settings Under Mattress Bed Sensor The Emfit under mattress bed sensor model L 4060SL is a quasi piezoelectric dyn...

Страница 4: ...tion at SW1 disabled device is always on 7 Notification sound volume setting see below Notification sound volume setting see below 8 Notification sound volume setting see below Notification sound volume setting see below Setting the Notification Sound Volume The notification sound is useful only when caregiver is in close proximity to the control unit The volume can be adjusted at 4 levels no soun...

Страница 5: ...ND SENSOR IS PREFERRED SET THE DIP SWITCH 4 UP THE SENSOR MUST ALWAYS BE PLACED BENEATH A MATTRESS OR MATTRESS PAD THE SENSOR MUST NOT BE IN DIRECT CONTACT WITH A PERSON THE EMFIT BED SENSOR IS DESIGNED TO WITHSTAND USE BELOW THE MATTRESS AGAINST A HARD BED BOTTOM FOR 4 5 YEARS OF USE MANUFACTURER RECOMMENDS REPLACING IT AFTER 6 YEARS FROM MANUFACTURING DATE OR AFTER 5 YEARS OF USE WHICHEVER COMES...

Страница 6: ...ONNECTED TO A SYSTEM SAFETY VOLTAGE INPUT WITH MAX VOLTAGE BELOW 25V AC 60V DC WHERE BOTH POLES HAVE BEEN SEPARATED FROM THE ELECTRICAL NETWORK SO CALLED FLOATING APPLIED PART MAX LOAD CURRENT 100 mA 7 INSTALLING THE WALL MOUNT AND DEVICE Connect the sensor wire plug to the control unit X3 marked connector If you need to connect the dry contact output to another system plug the suitable connection...

Страница 7: ... EMFIT EPILEPTIC TONIC CLONIC SEIZURE ALARM IS DESIGNED AND TESTED TO BE USED ONLY WITH GLOBTEK INC POWER SUPPLY MODEL NO GTM41060 1505 AND P N WR9QA3000LCP N MNK ANY OTHER TYPE OF AC ADAPTER MAY AFFECT THE PRODUCT S SAFETY The power supply is equipped with a blue indicator light when the light is on the power supply is in use If the light is off and the power supply is connected to the mains outl...

Страница 8: ...WITCH BY PASS SWITCH SW1 As a Reset Switch When the notification sound is activated it can be silenced by briefly pressing the the SW1 switch see picture A beep beep sound is also heard In case of faster continued movements when the faster movements have stopped the sound will also stop In case of no movement notification when person returns to the bed the sound will stop after about 40 seconds SW...

Страница 9: ...ing forward via the other system the device is connected to Check that the connection cable is connected correctly and it is in good condition Often the wire uses at both ends same connector but wiring order is not the same Make sure it is connected the right way Notification sound does not operate Check the volume level False notifications without reason Check the condition positioning and connec...

Страница 10: ...NTROL UNIT AND SENSOR DRY EXPOSURE TO EXCESSIVE MOISTURE CAN CAUSE IT TO MALFUNCTION THE PRODUCT FULLFILLS THE REQUIREMENTS OF THE EMC DIRECTIVE FOR MEDICAL DEVICES IT DOES NOT CAUSE ANY ELECTROMAGNETIC DISTURBANCE IN NORMAL WORKING CONDITIONS THE PRODUCT CAN BE STACKED OR PLACED NEAR OTHER PRODUCTS OR DEVICES AS LONG AS MECHANICAL VIBRATION IS NOT PRESENT ALWAYS CHECK THE FUNCTION OF THE PRODUCT ...

Страница 11: ...t with accessories peripheral equipment and other products of a type condition and standard other than prescribed by Emfit iv repair or attempted repair by persons who are not Emfit employees v adjustments or adaptations without Emfit s prior written consent including upgrading the product beyond specifications or features described in the instruction manual or modifications to the product to conf...

Страница 12: ...osition rotary switch for sensitivity adjustment Light indicators 3 LEDs green blue and red Notification delays Faster movements threshold delay adjustable 10 13 16 or 20 sec no movement notification threshold delay about 5 seconds Mounting On the wall with included bracket or on the table Dimensions mm 96 x 127 x 34 mm Weight 110 g Color White Enclosure protection IP20 Case Plastic Bed Sensor Mod...

Страница 13: ...liance level Electromagnetic environment Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air IEC 60601 1 2 test level Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines IEC 60601 1 2 test level Mains ...

Страница 14: ...requency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio ce...

Страница 15: ...tres m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 7...

Отзывы: