4
Table of DIP Switch Settings
Switch #
OFF (down)
ON (up)
#1
Movement continuation time delay setting, see next
chapter after volume settings
Movement continuation time delay setting, see next
chapter after volume settings
#2
Movement continuation time delay setting, see next
chapter after volume settings
Movement continuation time delay setting, see next
chapter after volume settings
#3
Movement frequency threshold 3 Hz (default)
Movement frequency threshold lowered to 2 Hz. Use
only if device does not notice person’s movements
(that the caregiver should be notified of) at OFF
position.
#4
Normal sensitivity for noticing micro movements;
use when sensor is installed below foam mattress or
mattress pad
Increased sensitivity for noticing micro movements.
Recommended use only when sensor needs to be
installed below spring mattress (no mattress pad in
use). NOTE! Consumes more power and shortens
battery life! Blued LED light blinks slightly faster.
#5
“No movements” notification is enabled
“No movements” notification is disabled
#6
Power switch function at SW1 is enabled (device
can be shut down and turned on by pressing SW1
for about 3 seconds)
Power switch function at SW1 disabled (device is
always on).
#7
Notification sound volume setting, see below
Notification sound volume setting, see below
#8
Notification sound volume setting, see below
Notification sound volume setting, see below
Setting the Notification Sound Volume
The notification sound is useful only when caregiver is in close proximity to the control unit. The volume can be adjusted at 4
levels: no sound, quiet, loud, and very loud.
Volume level
Switch # 7
Switch #8
Very Loud
OFF (down)
OFF (down)
Loud
ON (up)
OFF (down)
Quiet
OFF (down)
ON (up)
No sound / mute
ON (up)
ON (up)
The sound stops when both the reset switch is pressed and when the event is over. In case of person’s absence from the bed,
the sound will stop once person has returned or reset is pressed.
Setting the Movement Notifications
C
ontinued Faster Movements Notification Time Delay Setting
This notification is triggered when the device senses movements in frequency range between 3 Hz – 20 Hz (or 2 – 20 Hz if
switch #3 is ON) that continue longer than the preset delay. The alarm delay can be set at 10, 13, 16 or 20 seconds.
Set the preferred delay according to the following table. If false alarms become a problem, extend the delay.
Movements to continue
to trigger a notification
Switch # 1
Switch #2
10 seconds
ON(up)
ON (up)
13 seconds (default)
OFF (down)
OFF (down)
16 seconds
ON (up)
OFF (down)
20 seconds
OFF (down)
ON (up)