Emfit D-1090-2G Скачать руководство пользователя страница 14

 

 

  

14

 
 
 

Electromagnetic immunity 

 

 

Immunity test

 

 

IEC 60601 test 
level

 

 

Compliance 
level

 

 

Electromagnetic environment 

– 

guidance

 

 

 
 
 
 
 
 
 
 
 
 
Conducted RF 
IEC 61000-4-6 
 
 
 
 
 
 
 
Radiated RF 
IEC 61000-4-3

 

 

 
 
 
 
 
 
 
 
 
 
3 Vrms 
150 kHz to 80 MHz 
 
 
 
 
 
 
 
3 V/m 
80 MHz to 2,5 GHz 

 

 

 
 
 
 
 
 
 
 
 
 
 [

V

1

] = 3  V  

150 kHz to 80 MHz 

 

 
 
        
 
 
 
[

E

1

] = 3 V/m

 

80 MHz to 2,5 GHz

 

Portable and mobile RF communications equipment 
should be used no closer to any part of the 
The Emfit Nocturnal Epileptic Tonic-Clonic Seizure Monitor , 
including cables, than the recommended separation distance 
calculated from the equation applicable to the frequency of the 
transmitter. 
 

Recommended separation distance 

 

 

 
 
where 

is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 
manufacturer and 

is the recommended separation 

distance in metres (m). 
 
Field strengths from fixed RF transmitters, as determined 
by an electromagnetic site survey,

  

should be 

less than the compliance level in each frequency 
range.

 

 

Interference may occur in the vicinity of equipment 
marked with the following symbol:

 

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the [

EQUIPMENT 

or 

SYSTEM

] is 

used exceeds the applicable RF compliance level above, the [

EQUIPMENT 

or 

SYSTEM

] should be observed to verify 

normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting 
or relocating the [

EQUIPMENT 

or 

SYSTEM

]. 

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [

V

1

] V/m.

 

 
 
 
 

Содержание D-1090-2G

Страница 1: ...L CARE OR THAT WOULD LIKELY LEAD TO A POTENTIALLY LIFE THREATENING SITUATION Applicable model numbers Control Unit D 1090 2G version t43 v 1 0 10 50Hz or newer Under Mattress Bed Sensor L 4060SL Date June 27th 2008 V2 2 Emfi Emfit and Emfit logo are either registered trademarks or trademarks of Emfit Ltd in EU USA Japan and or other countries Copyright 2002 2008 Emfit Ltd All rights reserved Paten...

Страница 2: ...LECTROMAGNETIC CONDITIONS 13 21 MANUFACTURER S CONTACT INFORMATION 15 DANGER NOTES THE FOLLOWING NOTES ARE PROVIDED BOTH FOR YOUR PERSONAL SAFETY AND TO PROTECT THE DESCRIBED PRODUCT OR ATTACHED DEVICES FROM DAMAGE SAFETY NOTES AND WARNINGS FOR THE PREVENTION OF DANGER TO THE LIVES AND HEALTH OF USERS OR MAINTENANCE PERSONNEL AND OR FOR THE AVOIDANCE OF DAMAGE TO PROPERTY ARE EMPHASISED IN THESE I...

Страница 3: ...tems for transferring the notification via dry contact opto coupler A reset switch for turning off the notification sound is to the right of the sensor input marked as SW1 Inside the device there is a place for 2 pcs AA size 1 5 V batteries and mini switches DIP and rotary for program settings Under Mattress Bed Sensor The Emfit under mattress bed sensor model L 4060SL is a quasi piezoelectric dyn...

Страница 4: ...tion at SW1 disabled device is always on 7 Notification sound volume setting see below Notification sound volume setting see below 8 Notification sound volume setting see below Notification sound volume setting see below Setting the Notification Sound Volume The notification sound is useful only when caregiver is in close proximity to the control unit The volume can be adjusted at 4 levels no soun...

Страница 5: ...ND SENSOR IS PREFERRED SET THE DIP SWITCH 4 UP THE SENSOR MUST ALWAYS BE PLACED BENEATH A MATTRESS OR MATTRESS PAD THE SENSOR MUST NOT BE IN DIRECT CONTACT WITH A PERSON THE EMFIT BED SENSOR IS DESIGNED TO WITHSTAND USE BELOW THE MATTRESS AGAINST A HARD BED BOTTOM FOR 4 5 YEARS OF USE MANUFACTURER RECOMMENDS REPLACING IT AFTER 6 YEARS FROM MANUFACTURING DATE OR AFTER 5 YEARS OF USE WHICHEVER COMES...

Страница 6: ...ONNECTED TO A SYSTEM SAFETY VOLTAGE INPUT WITH MAX VOLTAGE BELOW 25V AC 60V DC WHERE BOTH POLES HAVE BEEN SEPARATED FROM THE ELECTRICAL NETWORK SO CALLED FLOATING APPLIED PART MAX LOAD CURRENT 100 mA 7 INSTALLING THE WALL MOUNT AND DEVICE Connect the sensor wire plug to the control unit X3 marked connector If you need to connect the dry contact output to another system plug the suitable connection...

Страница 7: ... EMFIT EPILEPTIC TONIC CLONIC SEIZURE ALARM IS DESIGNED AND TESTED TO BE USED ONLY WITH GLOBTEK INC POWER SUPPLY MODEL NO GTM41060 1505 AND P N WR9QA3000LCP N MNK ANY OTHER TYPE OF AC ADAPTER MAY AFFECT THE PRODUCT S SAFETY The power supply is equipped with a blue indicator light when the light is on the power supply is in use If the light is off and the power supply is connected to the mains outl...

Страница 8: ...WITCH BY PASS SWITCH SW1 As a Reset Switch When the notification sound is activated it can be silenced by briefly pressing the the SW1 switch see picture A beep beep sound is also heard In case of faster continued movements when the faster movements have stopped the sound will also stop In case of no movement notification when person returns to the bed the sound will stop after about 40 seconds SW...

Страница 9: ...ing forward via the other system the device is connected to Check that the connection cable is connected correctly and it is in good condition Often the wire uses at both ends same connector but wiring order is not the same Make sure it is connected the right way Notification sound does not operate Check the volume level False notifications without reason Check the condition positioning and connec...

Страница 10: ...NTROL UNIT AND SENSOR DRY EXPOSURE TO EXCESSIVE MOISTURE CAN CAUSE IT TO MALFUNCTION THE PRODUCT FULLFILLS THE REQUIREMENTS OF THE EMC DIRECTIVE FOR MEDICAL DEVICES IT DOES NOT CAUSE ANY ELECTROMAGNETIC DISTURBANCE IN NORMAL WORKING CONDITIONS THE PRODUCT CAN BE STACKED OR PLACED NEAR OTHER PRODUCTS OR DEVICES AS LONG AS MECHANICAL VIBRATION IS NOT PRESENT ALWAYS CHECK THE FUNCTION OF THE PRODUCT ...

Страница 11: ...t with accessories peripheral equipment and other products of a type condition and standard other than prescribed by Emfit iv repair or attempted repair by persons who are not Emfit employees v adjustments or adaptations without Emfit s prior written consent including upgrading the product beyond specifications or features described in the instruction manual or modifications to the product to conf...

Страница 12: ...osition rotary switch for sensitivity adjustment Light indicators 3 LEDs green blue and red Notification delays Faster movements threshold delay adjustable 10 13 16 or 20 sec no movement notification threshold delay about 5 seconds Mounting On the wall with included bracket or on the table Dimensions mm 96 x 127 x 34 mm Weight 110 g Color White Enclosure protection IP20 Case Plastic Bed Sensor Mod...

Страница 13: ...liance level Electromagnetic environment Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air IEC 60601 1 2 test level Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines IEC 60601 1 2 test level Mains ...

Страница 14: ...requency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio ce...

Страница 15: ...tres m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 7...

Отзывы: