125
Reference Manual
00809-0100-4021, Rev GC
Section 3: HART commissioning
April 2014
3.15.3
Transmitter-sensor matching
The 3144P accepts Callendar-Van Dusen constants from a calibrated RTD schedule and
generates a special custom curve to match that specific sensor resistance vs. temperature
performance. Matching the specific sensor curve with the transmitter significantly enhances the
temperature measurement accuracy. See the comparison below:
Problem description:
Depending on the process being measured, a certain amount of
accuracy may be needed from the sensor.
Our solution:
A more precise compensation for RTD inaccuracies is provided by Transmit-
ter-Sensor Matching using the transmitter's factory programmed Callendar-Van Dusen (CVD)
equation. This equation describes the relationship between resistance and temperature of
platinum resistance thermometers (RTDs). The matching process allows the user to enter the
four sensor specific Callendar-Van Dusen (CVD) constants into the transmitter. The transmitter
uses these sensor-specific constants in solving the CVD equation to match the transmitter to
that specific sensor thus providing outstanding accuracy. See comparison below:
Take away:
“Transmitter-Sensor Matching customizes sensor curves to minimize sensor
inaccuracy”
Note
In order to use this diagnostic, the RTD must be set as type
Cal VanDusen
.
HART 7 Fast Keys
Sensor 1 - 2, 2, 1, 11
HART 7 Fast Keys
Sensor 1 - 2, 2, 2, 11
System accuracy comparison at 150 °C using a PT 100 (
=0.00385)
RTD with a span of 0 to 200 °C
Standard RTD
Matched RTD
3144P
±0.08 °C
3144P
±0.08 °C
Standard RTD
±1.05 °C
Matched RTD
±0.18 °C
Total System
(1)
(1) Calculated using root-summed-squared (RSS) statistical method.
±1.05 °C
Total System
±0.21 °C
System accuracy comparison at 150 °C using a PT 100 (
=0.00385)
RTD with a span of 0 to 200 °C
Standard RTD
Matched RTD
3144P
±0.08 °C
3144P
±0.08 °C
Standard RTD
±1.05 °C
Matched RTD
±0.18 °C
Total System
(1)
(1) Calculated using root-summed-squared (RSS) statistical method.
±1.05 °C
Total System
±0.21 °C
Содержание Rosemount
Страница 1: ...Reference Manual 00809 0100 4021 Rev GC April 2014 Rosemount 3144P Temperature Transmitter ...
Страница 2: ......
Страница 4: ...iv Reference Manual 00809 0100 4021 Rev GC Title Page April 2014 ...
Страница 10: ...x Reference Manual 00809 0100 4021 Rev GC Table of Contents April 2014 Table of Contents ...
Страница 38: ...28 Reference Manual 00809 0100 4021 Rev GC Section 2 Installation April 2014 Installation ...
Страница 188: ...178 Reference Manual 00809 0100 4021 Rev GC Section 5 Maintenance April 2014 Maintenance ...
Страница 233: ...227 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 234: ...228 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 235: ...229 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 237: ...231 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 238: ...232 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 244: ...238 Reference Manual 00809 0100 4021 Rev GC Appendix B Product Certifications April 2014 Product Certifications ...
Страница 250: ...Index 244 Reference Manual 00809 0100 4021 Rev GC Index April 2014 Index ...
Страница 251: ......