Figure 4-4: Calibration methods, simplified
4.8.1
O
2
calibration with Rosemount Xi
For systems with configuration 1, shown in
, use the following procedure to
perform a calibration using the Rosemount Xi. If necessary, use the Rosemount Xi menu
tree in for reference.
NOTICE
To select a menu item, use the up and down arrow keys to scroll to the menu item and
press the right arrow key to select the menu item. To return to a preceding menu, press
the left arrow key.
1. From the Main Menu, select SYSTEM to access the Rosemount Xi System menu.
2. From the Rosemount Xi SYSTEM menu, scroll down and select DETAILED SETUP.
3. Select menu item 4, CAL SETUP, to input the cal gas and flow times.
4. Return to the SYSTEM menu and select the last menu item, CALIBRATION, to access
the CALIBRATION menu.
5. From the O
2
CALIBRATION options, select Start Cal to start the O
2
calibration
procedure.
WARNING
Failure to remove the Rosemount Xi from automatic control loops prior to
performing this procedure may result in a dangerous operating condition.
Reference Manual
Startup and operation
00809-0100-4889
October 2022