Set the calibration target either to a known value of the gas concentration or to zero. Then
set the desired number of cycles over which the calibration factor is determined
(Rosemount
™
recommends a minimum of 30). Now run the calibration, and the display will
show a countdown during its execution. When the calibration is complete, the new
Cal
Factor
is displayed for about five seconds, and the instrument exits the calibration routine.
Note
You must run the calibration routine during commissioning; otherwise, the instrument will not be
able to calculate the true level of gas within the duct.
Do not run the calibration routine unless reasonable conditions exist in the duct. If it is not
the initial calibration, record the
Cal Factor
from the
Parameters
option before running the
calibration.
To aid configuration and to record any subsequent changes to the operating parameters,
lists all of the options available and can be used as a record of the operating
parameters.
Instrument Settings
Table 3-1:
Parameter
Config
Averages
Seconds
Minutes
Hours
Days
Output
0 or 4 mA base
Units
Average
Configuration and startup
48
Rosemount CCO 5500
Содержание Rosemount CCO 5500
Страница 1: ...Reference Manual 00809 0100 4550 Rev AA June 2018 Rosemount CCO 5500 Carbon Monoxide CO Analyzer ...
Страница 4: ......
Страница 8: ...Contents iv Rosemount CCO 5500 ...
Страница 28: ...System Wiring Diagram Figure 2 7 Install 20 Rosemount CCO 5500 ...
Страница 32: ...Install 24 Rosemount CCO 5500 ...
Страница 88: ...Operation 80 Rosemount CCO 5500 ...
Страница 98: ...Maintenance 90 Rosemount CCO 5500 ...
Страница 108: ...Troubleshooting 100 Rosemount CCO 5500 ...
Страница 110: ...Spare parts 102 Rosemount CCO 5500 ...
Страница 122: ...Return equipment to the factory 114 Rosemount CCO 5500 ...
Страница 123: ...Return equipment to the factory Reference Manual 115 ...