Instruction Manual
IM-105-4000, Rev 1.3
October 2008
3-3
OPM 4000
Verifying Instrument
Operation and
Configuration
Before calibrating:
• Beam alignment procedure must be completed as outlined in Section 2:
• Apply power to both the control unit and the sensors for a minimum of
30 minutes before any adjustments are attempted.
• Observe the front panel LEDs and confirm the red operate and the
green RUN LEDs are on.
• Press the ACK button to confirm NO ALM'S message is displayed.
• Press the DIS button until OPACITY % is displayed.
EDIT KEYS
All data entry, editing and diagnostics are accomplished with four EDIT keys
behind the hinged cover (Figure 3-2).
QUICK MENU
1. Open the hinged cover and press the EDIT (ESC) button [the red EDIT
(ESC) LED will light]. The message display will read QUICK MENU.
2. Press the ENTER key then the EDIT
▼
key to scroll through the listing
shown below.
• Alarm Set Point
In % opacity
• Alarm Time Delay
In seconds
• Auto Cal Cycle Time In hours
• Zero Cal Value
In % opacity
• Span Cal Value
In % opacity
• OPLR Read Only
XX.XX
DISPLAYING THE
FACTORY SET OPLR
1. Open the hinged cover and press the EDIT (ESC) button [the red EDIT
(ESC) LED will light]. The message display will read QUICK MENU.
2. Press the ENTER key then the EDIT
▼
key or hold and scroll down
until the display reads "OPLR Read Only".
3. Press the ENTER key; the OPLR is displayed.
The OPLR is a read only block and cannot change the OPLR setting. It
is intended for a convenient place to view the OPLR. The OPLR is
factory set and cannot be changed in the field. If an OPLR change is
required, contact the factory for return authorization number and
shipping instructions.
Содержание Rosemount Analytical OPM 4000
Страница 2: ......
Страница 6: ......
Страница 10: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 TOC 4 ...
Страница 14: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 iv ...
Страница 38: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 16 Figure 2 12 Typical System Wiring ...
Страница 42: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 2 20 ...
Страница 52: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 4 6 ...
Страница 58: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 5 6 ...
Страница 64: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 6 6 ...
Страница 100: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 A 24 ...
Страница 102: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 B 2 ...
Страница 104: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 C 2 ...
Страница 110: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 6 Figure D 5 Retrofit System Wiring Diagram ...
Страница 114: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 OPM 4000 D 10 ...
Страница 116: ...Instruction Manual IM 105 4000 Rev 1 3 October 2008 Index 2 OPM 4000 ...