![Emerson Rosemount 644 with HART Protocol Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/emerson/rosemount-644-with-hart-protocol/rosemount-644-with-hart-protocol_reference-manual_104732023.webp)
15
Reference Manual
00809-0200-4728, Rev SA
Configuration
July 2018
Configuration
Note
The list of choices available for Units after the primary menu is dependent on your Sensor configuration
settings.
2.7
Configure dual sensor options
Dual-sensor configuration deals with the functions that can be used with a transmitter ordered with Dual
Sensor inputs. In the Rosemount 644 Transmitter these functions include:
Differential temperature
Average temperature
Hot Backup
™
and sensor drift alert diagnostics (requires option code DC)
–
First good temperature (requires options S and DC, or options D and DC)
2.7.1
Differential temperature configuration
The Rosemount 644 Transmitter ordered and configured for dual-sensors can accept any two inputs then
display the differential temperature between them. Use the following procedures to configure the
transmitter to measure differential temperature.
Note
This procedure assumes the differential temperature is a calculated output of the device but does not
re-assign it as the primary variable. If it desired for Differential to be the transmitter’s primary variable
see
“Mapping the HART variables” on page 11
to set it to PV.
Field Communicator
From the
HOME
screen, enter the Fast Key sequence.
AMS Device Manager
1. Right click on the device and select
Configure
.
2. In the left navigation pane choose
Manual Setup
.
3. On the
Calculated Output Tab
find the
Differential Temperature
group box.
4. Select Units and Damping settings then select
Apply
when complete.
Device Dashboard Fast Keys
2, 2, 3, 1
Содержание Rosemount 644 with HART Protocol
Страница 2: ......
Страница 8: ...vi Reference Manual 00809 0200 4728 Rev SA Title Page July 2018 Title Page ...
Страница 48: ...40 Reference Manual 00809 0200 4728 Rev SA Configuration July 2018 Configuration ...
Страница 62: ...54 Reference Manual 00809 0200 4728 Rev SA Installation July 2018 Installation ...
Страница 72: ...64 Reference Manual 00809 0200 4728 Rev SA Operation and Maintenance July 2018 Operation and Maintenance ...
Страница 96: ...Reference Data July 2018 Reference Manual 00809 0200 4728 Rev SA Specifications and Reference Data 88 ...
Страница 111: ......