![Emerson Rosemount 644 FOUNDATION Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/emerson/rosemount-644-foundation/rosemount-644-foundation_reference-manual_104730032.webp)
26
Reference Manual
00809-0400-4728, Rev CA
Configuration
September 2018
Configuration
2.5.6
LCD transducer block
This section describes error conditions found in the LCD Transducer Block. Read
and to determine the appropriate corrective action.
Table 2-8. Resource Block BLOCK_ERR Messages
Condition name and description
Other
Device Needs Maintenance Now
Memory Failure: A memory failure has occurred in FLASH, RAM, or EEPROM memory.
Lost NV Data: Non-volatile data that is stored in non-volatile memory has been lost.
Device Needs Maintenance Now.
Out of Service: The actual mode is out of service.
Table 2-9. Resource Block SUMMARY_STATUS Messages
Condition name
No repair needed
Repairable
Call Service Center
Table 2-10. Resource Block RB.DETAILED_STATUS
RB.DETAILED_STATUS
Description
Sensor Transducer block error
Active when any SENSOR_DETAILED_STAUS bit is on.
Manufacturing Block integrity error
The manufacturing block size, revision, or checksum is wrong.
Hardware/software incompatible
Verify the manufacturing block revision and the hardware revision
are correct/compatible with the software revision.
Non-volatile memory integrity error
Invalid checksum on a block of NV data.
ROM integrity error
Invalid application code checksum.
Lost deferred NV data
Device has been power-cycled while non-volatile writes were being
deferred to prevent premature memory failure, the write operations
have been deferred.
NV Writes Deferred
A high number of writes has been detected to non-volatile memory.
To prevent premature failure, the write operations have been
deferred.
Содержание Rosemount 644 FOUNDATION
Страница 2: ......
Страница 6: ...4 Reference Manual 00809 0100 4728 Rev CA Title Page September 2018 Title Page ...
Страница 10: ...4 Reference Manual 00809 0400 4728 Rev CA Introduction September 2018 Introduction ...
Страница 34: ...28 Reference Manual 00809 0400 4728 Rev CA Configuration September 2018 Configuration ...
Страница 48: ...Reference Data September 2018 Reference Manual 00809 0400 4728 Rev CA Reference Data 42 ...
Страница 63: ......