It is assumed that the personnel installing, configuring, and operating the system have the
knowledge equal or greater than that of a qualified Instrument Technician familiar with
safety-related systems, process control applications, and general instrument use.
Note
The Rosemount 5300 Level Transmitter is not safety-rated during maintenance work,
configuration changes, multidrop, loop test, or other activity that affects the safety
function. Alternative means should be used to ensure process safety during such activities.
8.6
Configuring in SIS applications
Use a HART-compliant master, such as Rosemount Radar Master or a handheld
communicator, to communicate with and verify configuration of the Rosemount 5300
Level Transmitter. A full review of configuration methods is available in
.
These instructions are applicable to the Rosemount 5300 Level Transmitter safety-
certified options with any differences noted.
8.6.1
Damping
User adjusted damping will affect the transmitter’s ability to respond to process changes.
Therefore, the
damping values
+
response time
should not exceed the loop requirements.
For further information on damping, see
8.6.2
Alarm and saturation levels
DCS or safety logic solver should be configured to handle both High alarm and Low alarm.
In addition, the transmitter must be configured for High or Low alarm (see
identifies the alarm levels available and their operation values
(8)
In certain cases, the transmitter does not go into the user defined alarm state. For example, in case of a short circuit, the
transmitter goes into High Alarm state even if Low Alarm has been configured.
Reference Manual
Safety Instrumented Systems (4-20 mA only)
00809-0100-4530
April 2019
Reference Manual
249
Содержание Rosemount 5300 Series
Страница 1: ...Reference Manual 00809 0100 4530 Rev FA April 2019 Rosemount 5300 Level Transmitter Guided Wave Radar ...
Страница 26: ...Transmitter overview Reference Manual April 2019 00809 0100 4530 26 Rosemount 5300 Level Transmitter ...
Страница 112: ...Electrical installation Reference Manual April 2019 00809 0100 4530 112 Rosemount 5300 Level Transmitter ...
Страница 158: ...Configuration Reference Manual April 2019 00809 0100 4530 158 Rosemount 5300 Level Transmitter ...
Страница 168: ...Operation Reference Manual April 2019 00809 0100 4530 168 Rosemount 5300 Level Transmitter ...
Страница 244: ...Service and troubleshooting Reference Manual April 2019 00809 0100 4530 244 Rosemount 5300 Level Transmitter ...
Страница 380: ...Product certifications Reference Manual April 2019 00809 0100 4530 380 Rosemount 5300 Level Transmitter ...
Страница 396: ...Configuration parameters Reference Manual April 2019 00809 0100 4530 396 Rosemount 5300 Level Transmitter ...
Страница 487: ...Reference Manual 00809 0100 4530 April 2019 Reference Manual 487 ...