Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
3-24
Slope
Refer to Figure 3-32. Use this screen to examine the slope calculated from
current calibration check. The slope is the amount of cell voltage generated
for a given O
2
value. After each calibration check, record the slope over the
life of the probe. Tracking the slope will indicate if the probe is degrading.
Press NEXT to access the CONSTANT screen.
Constant
Refer to Figure 3-33. Use this screen to examine the cell zero constant
calculated from the current calibration check. The constant represents the
voltage generated by the cell when no difference exists between the amount
of O
2
on the reference and process sides of the cell. Note this value for
comparison against future calibration checks. Press RESET or EXIT to return
to the PROCESS DISPLAY screen.
Figure 3-32. Slope
Figure 3-33. Constant
Содержание Rosemount 5081FG
Страница 2: ......
Страница 6: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG TOC 4 ...
Страница 20: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 1 10 ...
Страница 23: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 3 Figure 2 1 Probe Installation Details ...
Страница 26: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 6 Figure 2 4 Horizontal Probe Installation ...
Страница 29: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 9 Figure 2 7 Pipe Mounting Dimensions ...
Страница 36: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 16 ...
Страница 41: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 5 Figure 3 5 Transmitter Menu Tree ...
Страница 53: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 17 Figure 3 20 Show Fault ...
Страница 81: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 5 Figure 6 2 Oxygen Probe Terminal Block ...
Страница 82: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 6 ...
Страница 94: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG A 8 ...
Страница 96: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Страница 97: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG EC Declaration of Conformity ...
Страница 98: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Страница 100: ......
Страница 103: ...Index 2 ...