Reference Manual
00809-0100-4882, Rev AB
May 2019
Rosemount 5081FG
1-8
Electronics
Electronics Enclosure
TYPE 4X (IP65), weatherproof, and
corrosion-resistant
Materials of Construction
Low copper aluminum
Ambient Temperature Limits
-20° to 65°C (-4° to 149°F)
Relative Humidity
95% with covers sealed
Power Supply and Load Requirements See Figure 1-4
Inputs (from O
2
Probe)
Two wires - O2 signal
Two wires - type B thermocouple
Output
One 4-20 mA signal with superimposed digital HART
signal
Hazardous Area Certification
See Section 8.
Power Transient Protection
IEC 801-4
Hazardous Area Certification
ATEX EEx ia IIC Class 1, Div 1, Group A, B, C, D
Figure 1-4. Power Supply and
Load Requirements
Shipping Weight
10 lbs (4,5 kg)
Infrared Remote Control
Power Requirements
Three AAA atteries
Содержание Rosemount 5081FG
Страница 2: ......
Страница 6: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG TOC 4 ...
Страница 20: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 1 10 ...
Страница 23: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 3 Figure 2 1 Probe Installation Details ...
Страница 26: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 6 Figure 2 4 Horizontal Probe Installation ...
Страница 29: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 9 Figure 2 7 Pipe Mounting Dimensions ...
Страница 36: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 2 16 ...
Страница 41: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 5 Figure 3 5 Transmitter Menu Tree ...
Страница 53: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 3 17 Figure 3 20 Show Fault ...
Страница 81: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 5 Figure 6 2 Oxygen Probe Terminal Block ...
Страница 82: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG 6 6 ...
Страница 94: ...Reference Manual 00809 0100 4882 Rev AB May 2019 Rosemount 5081FG A 8 ...
Страница 96: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Страница 97: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG EC Declaration of Conformity ...
Страница 98: ...Reference Manual 00809 0100 4882 Rev AB June 2019 Rosemount 5081FG B 2 ...
Страница 100: ......
Страница 103: ...Index 2 ...