Off-line and On-line Operation
The Model 375 Communicator features off-line and on-line communications. On-line means the communicator is
connected to the transmitter in the usual fashion. While the communicator is on line, the operator can view meas-
urement data, change program settings, and read diagnostic messages. Off-line means the communicator is not
connected to the transmitter. When the communicator is off line, the operator can still program settings into the
communicator. Later, after the communicator has been connected to a transmitter, the operator can transfer the
programmed settings to the transmitter. Off-line operation permits settings common to several transmitters to be
easily stored in all of them.
9.3.2
Making HART related settings from the keypad
9.3.3
Menu Tree
The menu tree for the Model 375 HART communicator is on the following page. The menu tree for the Model 375
F
OUNDATION
Fieldbus communicator immediately follows.
Note:
Model 375 Communicator fully supports F
OUNDATION
Fieldbus on Model 5081-T-FF.
1.
Press MENU. The main menu screen appears. Choose
Program
.
2. Choose
>>
.
3. Choose
HART
.
4. To display the device ID, choose
DevID
. To change the polling address,
choose
PollAddrs
. To make burst mode settings, choose
Burst
. To
change the preamble count, choose
Preamble
.
MODEL 5081-T
SECTION 9.0
OPERATION WITH MODEL 375
Calibrate
Hold
P
Pr
ro
og
gr
ra
am
m
Display
Output
Temp
Measurement
>
>>
>
Security
H
HA
AR
RT
T
>>
D
De
ev
vI
ID
D
PollAddrs
Burst
Preamble