52
MODEL 5081-T
SECTION 7.0
PROGRAMMING
7.2.3 Procedure
1.
Press PROG on the remote controller. The
OutPut
submenu appears.
2.
Press ENTER. The screen displays the
4 MA
prompt. Use the arrow keys to change
the setting. Press ENTER to save.
3.
The screen displays the
20 MA
prompt. Use the arrow keys to change the setting.
Press ENTER to save.
4.
The screen displays the
HoLd
prompt. Use the arrow keys to change the setting to
the output desired when the transmitter is in hold. The range is 3.80 to 22.00 mA.
Entering 00.00 causes the transmitter to hold the output at the value it was when
placed in hold. The hold setting overrides the fault setting. Press ENTER to save.
5.
The screen displays the
FAULt
prompt. Use the arrow keys to change the setting to
the output desired when the transmitter detects a fault. The range is 3.80 to 22.00 mA.
Entering 00.00 causes the transmitter to hold the output at the value it was when the
fault occurred. Press ENTER to save.
6.
The screen displays the
dPn
prompt. Use the arrow keys to change the setting. The
range is 0 to 255. Press ENTER to save.
7.
The screen displays the
tESt
prompt. Use the arrow keys to enter the desired test cur-
rent. Press ENTER to start the test. Press EXIT to end the test.
8.
Press RESET to return to the process display.