MODEL 5081-T
SECTION 7.0
PROGRAMMING
7.7 RANGE
7.7.1 Purpose
This section provides the steps to select automatic ranging or a specific fixed range of measurement. Five specific con-
ductance ranges are selectable. Setting the Model 5081 to a fixed range reduces response time.
The following conductance ranges are available:
NOTE:
The selection between automatic ranging or a specified fixed range of measurement range can ONLY be done
using the IRC/Infrared Remote Controller. This selection cannot be done via HART or F
OUNDATION
Fieldbus hosts or
configurators.
7.7.2 Procedure
1.
Press
PROG
.
2.
Press
NEXT
five times.
S-PJU!!
(range) will appear.
3.
Press
ENTER
. The default setting
Auto
will appear. This indicates that Model 5081 is in auto ranging mode.
4.
To set a fixed conductance range, press the down arrow key
until the desired measurement range appears.
1400mS, 550mS, 200mS, 33mS
and
3mS
will appear when pressing the down arrow
successively.
5.
When the desired range is reached, press
ENTER
. This disables auto ranging and limits the measurement to the
selected range.
6.
Press
NEXT
to move to the next Programming menu item.
Setting
Measurement Range Over Range Warning Limit
Auto
0uS to 1400mS
None
1400mS
550mS to 1400mS
None
550mS
200mS to 550mS
570mS
200mS
33mS to 200mS
207mS
33mS
3000uS to 33mS
34mS
3mS
0uS to 3000uS
3400uS
1.
Press PROG on the remote controller.
2.
Press NEXT until the
dEFAULt
appears in the display. Press ENTER.
3.
Use
or
to toggle between
nO
and
yES
. With
yES
showing, press ENTER to return
to factory default settings.
PROGRAM
2-FUQSA
EXIT
ENTER
P9
PROGRAM
G12->[U
EXIT
NEXT
ENTER
7.8 DEFAULT
7.8.1 Purpose
This section describes how to erase ALL user-defined configuration settings and return the transmitter to factory default
settings. All custom curve values and settings will be deleted.
7.8.2 Procedure
57